20
PRS-812 Resistance Meter
7. Measurements Less than 10,000 ohms (<1.0x10
4
ohms):
a. When the resistance measurement is less than 1.0x10
4
ohms, test voltage will remain
at
<10V
, the resistance range will be adjusted automatically and a stable resistance
measurement will be obtained within 2.0 to 3.0 seconds.
b. Once a stable measurement is confirmed and displayed by the instrument,
HOLD
will
be energized in the LCD and the electrification timer will stop.
c. An LED corresponding to the measurement’s order of magnitude will be energized
Figure 10:
Comparing PRS-812 Display in Automatic Ohms Mode and Automatic Exponential Mode after the same
low resistance measurement.
8. Measurements from 1.0x10
4
to less than 1.0x10
6
ohms:
a.
When the resistance measurement is
greater than 1.0x10
4
ohms test voltage
will automatically be increased to
10
volts, and the resistance range will be
adjusted as necessary.
b.
A stable resistance measurement will be
obtained within 2.0 to 4.0 seconds.
c.
Once a stable measurement is
confirmed and displayed,
HOLD
will be
energized in the LCD and the
electrification timer will stop.
d.
An LED corresponding to the
measurement’s order of magnitude will
be energized
Measuring 958
in Auto Ohms Mode
Measuring 958
in Auto Exponential Mode
Analog Scale &
X100 Multiplier
<E03
LED
<E03
LED
RECord &
Ohms Range
Indicators
<10 Volts
Test Voltage
HOLD Indicates
Measurement
Is Complete
RECord &
Ohms Range
Indicators
<10 Volts
Test Voltage
HOLD Indicates
Measurement
Is Complete
Electrification Remains 2.0 – 3.0 sec.
on Measurements <1.0E+06 Ohms
Test Voltage Automatically shifts to
10 V @ 1.0E+04 Ohms
Note K
Indicator
E04
LED
Figure 11:
Test Voltage at 10 Volts from 1.0E+04 to
<1.0E+06 Ohms in Auto