![Prostat PRS-801 User Manual Download Page 35](http://html1.mh-extra.com/html/prostat/prs-801/prs-801_user-manual_1619194035.webp)
35
Rev. C / January 2012
PRS-812 Resistance Meter
PRS-812 Instrument Controls
FUNCTION/ MODE Toggles Through Six (6) Operation Modes:
MODE
DISPLAYED
UNITS
INDICA
-
TION
RESIST.
RANGE
TEST
VOLTS
TEST
FUNCTIONS
AUTOMATIC 1
[Default]
1.0EXX
IND Ω - TΩ
AUTO
AUTO.
AUTO.
AUTOMATIC RESISTANCE RANGE,
TEST VOLTS, ELECTRIFICATION,
DISPLAY HOLD
SEE TABLE NOTE #1
AUTOMATIC 2
OHMS
Ω - TΩ
AUTO
AUTO.
AUTO.
MANUAL 1
1.0EXX
IND Ω - TΩ
MANUAL
MAN
MAN. or
AUTO
TEST ONLY; NO DISPLAY HOLD
UL= UNDER DECADE LEVEL
OL= OVER DECADE LEVEL
SEE TABLE NOTE #2
MANUAL 2
OHMS
Ω - TΩ
MANUAL
MAN
MAN. or
AUTO
MANUAL/
AUTOMATIC 1
1.0EXX
IND Ω - TΩ
AUTO &
MANUAL
MAN. Start
AUTO
Run
AUTO.
MANUAL SETUP STARTING
DECADE, AUTO ADJUST RANGE,
TEST VOLTS, ELECTRIFICATION
DISPLAY HOLD
SEE TABLE NOTE #3
MANUAL/
AUTOMATIC 2
OHMS
Ω - TΩ
AUTO &
MANUAL
MAN. Start
AUTO
Run
AUTO.
TABLE NOTES:
#1 AUTOMATIC:
RESISTANCE RANGES IN AUTO 0.1
Ω
TO <2.0E+12
Ω
as follows
@<10V: 0.1 TO <1.0E+04
Ω
(0.1
Ω
- < 10K
Ω
)
@ 10V:
1.0E+04 TO <1.0E+06
Ω
(10K
Ω
- < 1M
Ω
)
@100V:
1.0E+06 TO <2.0E+12
Ω
(1M
Ω
- <200T
Ω
)
See Note Below
#2 MANUAL:
RESISTANCE RANGES IN MAN 0.1
Ω
TO <2.0E+12
Ω
as follows
@<10V: 0.1 TO <1.0E+05
Ω
(0.1
Ω
- <100K
Ω
)
@ 10V:
1.0E+03 TO <1.0E+09
Ω
(1K
Ω
- <1G
Ω
)
@100V:
2.0E+05 TO <2.0E+12
Ω
(200K
Ω
-< 200T
Ω
)
See Note Below
#3 AUTO-MANUAL: (Same as AUTOMATIC)
RESISTANCE RANGES IN AUTO-MANUAL 0.1
Ω
TO <2.0E+12
Ω
as follows
@<10V:
0.1 TO <1.0E+04
Ω
(0.1
Ω
- < 10K
Ω
)
@ 10V:
1.0E+04 TO <1.0E+06
Ω
(10K
Ω
- < 1M
Ω
)
@100V:
1.0E+06 TO <2.0E+12
Ω
(1M
Ω
- <200T
Ω
)
See Note Below
NOTE: It is capable of measuring up to 2.0x10
14
ohms with tolerance depending on procedures
and conditions.
Resistance
2 Triangular button Arrows: UP (↑) and DOWN (↓). Select Resistance
Range Select.
Range in single decades in Manual and Automatic/Manual modes
Test Volts
Manual Selection of <10, 10 or 100 volts in Manual Mode
Record/
Recall
Turns Memory Register ON if OFF. Provides access to all data in Memory Regis-
Summary of Contents for PRS-801
Page 1: ...PRS 812 RESISTANCE METER User Manual ...
Page 2: ......
Page 37: ...NOTES 37 Rev C January 2012 PRS 812 Resistance Meter ...
Page 38: ...NOTES 38 Rev C January 2012 PRS 812 Resistance Meter ...
Page 39: ......