16
Rev. C / January 2012
PRS-812 Resistance Meter
F. Properly dispose of any old batteries.
G. Position the new alkaline battery under the battery cover.
H. With your thumb, press the terminals in place one at a time.
I. Carefully wind any excess battery lead around the battery terminal below the cap. To avoid instru-
ment damagae, do not pull battery connection leads
J. Place the batteries back into position with the battery terminal covers facing the lead spurce
within the battery case shown in figure 4.
K. Carefully position the instrument battery cover back into place without force.
L. Replace the battery cover screws.
M. Slide battery buss switch to the
ON
position.
Following the above directions for battery replacement will insure that you do not damage the battery
covers or wires during this process.
NOTE
It is recommended to measure the voltage of each battery with a multimeter. If the voltage of
one battery measures at or below 6.75 volts, replace that battery. If both batteries measure at
or below 6.75 volts, replace both batteries.
V. Setup & Calibration
A. Setting up the PRS-812 for low resistance range calibration
1. Position Battery Buss Cut Off switch to OFF position
2. To install batteries, remove two Phillips locking screws and cover. Attach two Long-Life, Alkaline
9V batteries to the battery connection terminals.
3. Position batteries in compartment with power leads neatly positioned above battery connec-
tions. Carefully re-install battery cover and locking screws.
NOTE
Do Not change batteries with the battery buss
ON
. Always switch Battery Buss Cut Off to
OFF
when changing batteries. Should the
ON/OFF
button be depressed during battery change and
Battery Bus is
ON
the instrument may lock up and not function properly or be seriously dam-
aged. In this case, simply disconnect the batteries, then re-install with Battery Buss switch in
the
OFF
position.
B. Low Resistance Range (<10 Ohms) Calibration
1. Install the calibration shunt (PRS-801CC) across the Negative (-) and Positive (+) Lead Terminals
(Figure 5, below).