13
Rev. E / October 4, 2018
PGA-710B Autoanalysis System
a. The
Reset Voltage
starts the ming calcula on period.
b. Cuto
ff
Voltage
ends the ming calcula on period.
2. The Decay Window o
ff
ers two means to designate
Reset
and
Cuto
ff
Voltages
.
a. Standard
Reset
and
Cuto
ff
Voltages
are (See Figure 12):
1,000 Volts – 100 Volts
1,000 Volts – 50 Volts
1,000 Volts – 10 Volts
b. Check the Custom Box to enter any two voltages for evalua on. Figure 12.
F. Specialty Screens: Voltage Genera on Tests
Voltage genera on measurements may
take several forms. In many tests, the
movement is repe ous, much like a walk-
ing measurement. The Wizard screen at
the right (Figure 13) allows the operator to
es mate the number of cycles before the
measurements begin. The number can be
changed later as necessary.
However, it is the number of operator en-
tered cycles that drive the so ware’s ini al
algorithm analysis. Thus, it is important to
be able to de
fi
ne the number of walking
steps or test cycles so that automa c analy-
sis proceeds e
ff
ec vely.
G. Customize Charts & Create Templates
Figure 12: Standard Reset and Cuto
ff
Voltages (le ), or check the Custom Box (right) and Manually Enter
Figure 13: Enter the Number of es mated Voltage Gen-
era on Test Cycles