background image

 

 

4.2.7 Diode

 

Functions 

Testing conditions 

Diode test 

 

Forward DC current: approximately 1.2mA; Open-circuit 

voltage: approximately 3.2V.   

The display unit shows the approximate value of the diode's 

forward voltage drop.   

 

The buzzer beeps when the resistance is less than 30

   

Overload protection: 250V DC/AC 

4.2.8 DC current   

Measuring range 

Resolution 

Accuracy 

60mA 

0.01mA 

±(1.2% R 5 digits) 

600mA 

0.1mA 

10.00A 

10mA 

±(3% R 5 digits) 

Overload protection: 

Protective tube for mA measuring range (F600mA/250V) ;   

Protective tube for 10A measuring range (F10A/250V) . 

Maximal input voltage: mA position: 600mA DC/AC (valid value); 

10A position: 10A DC/AC (valid value) 

When the measured current is over 5A, the duration of continuous measurement 

shall not be over 10 seconds. The current measurement shall be carried out 1 

minute after the completion of previous measurement.   

4.2.9 AC current   

Measuring range 

Resolution 

Accuracy 

60mA 

0.01mA 

±(1.5% R 5 digits) 

600mA 

0.1mA 

10.00A 

10mA 

±(3% R 5 digits) 

Overload protection: 
Protective tube for mA measuring range (F600mA/250V) ;   
Protective tube for 10A measuring range (F10A/250V) . 
Maximal input voltage: mA position: 600mA DC/AC (valid value); 

10A position: 10A DC/AC (valid value) 

When the measured current is over 5A, the duration of continuous 
measurement shall not be over 10 seconds. The current measurement shall 
be carried out 1 minute after the completion of previous measurement.   
Frequency response: 40Hz-1kHz True RMS 
 

 

Summary of Contents for MT-1236

Page 1: ...MT 1236 3 5 6 True RMS Multimeter User s Manual 1st Edition 2020 2020 Copyright by Prokit s Industries Co Ltd...

Page 2: ...e null if the meter is not properly operated The meter must be operated with great care when working in the vicinity of an exposed conductor or bus line The meter is prohibited from being used in the...

Page 3: ...When appears the batteries must be replaced immediately The low level of a battery will result in incorrect reading on the meter which is likely to bring electric shock or personal injury to the opera...

Page 4: ...ee to it that the meter cannot be in service unless it is proved conforming When the meter is left idle for a long period the operator shall remove the battery and place it in a place free from high t...

Page 5: ...ata hold mode NCV Non contact AC voltage detection mode Automatic range measurement mode Unit of temperature Celsius Fahrenheit V mV V mV V the unit of volt Millivolt 1x10 3 or 0 001 volt A mA A mA Am...

Page 6: ...rn off the illumination function It turns off automatically with no operation for 30S 2 4 Description of input socket Input socket Description COM All public input terminals to be measured are connect...

Page 7: ...electric shock and or damage to the meter don t attempt to impose between any public terminal and ground any voltage whose valid value is over 600V for DC or AC current The meter provides DC AC voltag...

Page 8: ...the test lead to measure the electric resistance of the circuit to be measured 4 Read the measured electric resistance value on LCD screen Notes The measured value of the electric resistance of the c...

Page 9: ...s than 30 the beeper will sound continuously and the LED will be on 3 2 5 Capacitance measurement To avoid the meter or the measured equipment from damage do not attempt a capacitance measurement unle...

Page 10: ...test lead in black to COM input socket Connect the test lead in red to a mA input socket when the measured current is less than 600mA connect the test lead in red to a 10A input socket when the measu...

Page 11: ...frequencies and the indicator will be flashing Note 1 Voltage may still remain in the absence of any indication The operator shall not rely on non contact voltage detector to check the presence of vo...

Page 12: ...AAA battery Dimensions 148mm L 79mm W 48mm H Weight approximately 210g not included batteries or test leads 4 2 Precision indicator Accuracy reading digit The accuracy warranty will run for 1 year up...

Page 13: ...Accuracy 1 99 0 1 2 5 Reading 5 digits Input voltage range 200mV 10V AC valid value Overload protection 250V DC AC 4 2 5 Capacitor Measuring range Resolution Accuracy 60nF 0 01nF 4 Reading 5 digits 6...

Page 14: ...on 10A DC AC valid value When the measured current is over 5A the duration of continuous measurement shall not be over 10 seconds The current measurement shall be carried out 1 minute after the comple...

Page 15: ...f any abrading or chemical solvent The dirty or damp input socket may affect reading Steps for cleaning input sockets Disenable the meter and pull all test leads out of the input socket Clean up all d...

Page 16: ...replacement When fuse is blown replace with the same type of fuse 1 Turn off the power of the meter and take out the holster 2 Use a screwdriver to remove the screw of back cover 3 Take off the back...

Page 17: ...1 Pro sKit MT 1236 1 1 A B 60V DC 30V AC 3 1 5V AAA 600V...

Page 18: ...AC DC 600 V European Union 1 2 600V 250V mA F600mA 250V A F10A 250V 2 2 1 LED...

Page 19: ...2 2 NCV V mV V mV 1x10 3 0 001 A mA A mA 1x x10 3 0 001 MkHz Hz kHz 1x103 MHz 1x106 1000 mF F nF F mF 1x10 3 0 001 F 1x10 6 0 000001 nF 1x10 9 0 000000001 k M k 1000 M 1 000 000 2 3 HOLD...

Page 20: ...FUNC Hz Hz 15 3 30 2 4 COM V mA Hz mA 10A 10A 2 5 3 3 1 HOLD H H 15s 30s 30s 15 3 2 600V 600V 600 0mV 6 000V 60 00V 600V...

Page 21: ...V FUNC Hz DCV ACV COM V 600 0 6 000k 60 00k 600 0k 6 000M 60 00M FUNC Hz COM V OL FUNC Hz COM V OL 0 5V 0 8V FUNC Hz COM V 30 60 00nF 600 0nF 6 000 F 60 00 F 600 0 F 6mF 100mF...

Page 22: ...COM 250V Hz COM Hz FUNC Hz Hz 250V 60 00mA 600 0mA 10A mA A FUNC Hz COM 600mA mA 600mA 10A 10A OL FUNC Hz COM V NCV NCV...

Page 23: ...0 1 18 28 600V mA F600mA 250V A F 10A 250V 3 6000 counts OL 3 x 1 5V AAA 148 mm L 79mm W 48mm H 210g 4 2 18 28 80 600mV 0 1mV 0 5 5 6V 0 001V 60V 0 01V 600V 0 1V 10M 600V DC AC 600mV 0 1mV 1 4 6V 0 0...

Page 24: ...C 9 999Hz 0 001Hz 1 5 5 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999KHz 0 001KHz 99 99KHz 0 01KHz 999 9KHz 0 1KHz 9 999MHz 0 001MHz 200mV 10V AC 250V DC AC 1 99 0 1 2 5 5 200mV 10V AC 250V DC AC 60nF 0 01nF 4 5...

Page 25: ...250V DC AC 60mA 0 01mA 1 2 5 600mA 0 1mA 10 00A 0 01A 3 5 mA F600mA 250V 10A F10A 250V mA 600mA 10A 10A 5A 10 1 60mA 0 01mA 1 5 5 600mA 0 1mA 10 00A 0 01A 3 5 mA F600mA 250V 10A F10A 250V 40Hz 1kHz mA...

Page 26: ...5 2 F600mA 250V F10A 250V 3 x 1 5V AAA...

Page 27: ...MT 1236 3 5 6 1 Pro sKit MT 1236 1 1 A B 60Vdc 30Vac 3 1 5V AAA...

Page 28: ...600V AC DC CAT 600 V European Union 1 2 600V 250V mA F600mA 250V A F10A 250V...

Page 29: ...2 2 1 LED 2 2 NCV V mV V...

Page 30: ...001 A mA A mA 1x x10 3 0 001 MkHz Hz kHz 1x103 MHz 1x106 1000 mF F nF F mF 1x10 3 0 001 F 1x10 6 0 000001 nF 1x10 9 0 000000001 k M k 1000 M 1 000 000 2 3 HOLD FUNC Hz Hz 15 3 30 2 4 COM V mA Hz mA 10...

Page 31: ...3 1 HOLD H H 15s 30s 30s 15 3 2 600V 600V 600 0mV 6 000V 60 00V 600V V FUNC Hz COM V 600 0 6 000k 60 00k 200 0k 6 000M 60 00M FUNC Hz COM V...

Page 32: ...60M OL FUNC Hz COM V OL 0 5V 0 8V FUNC Hz 30 60 00nF 600 0nF 6 000 F 60 00 F 600 0 F 6mF 100mF COM...

Page 33: ...250V Hz COM Hz FUNC Hz Hz 250V 60 00mA 600 0mA 10 00A COM 600mA mA 600mA 10A 10A OL FUNC Hz COM V NCV NCV 4 4 1...

Page 34: ...0 70 RH 0 1 18 28 600V mA F600mA 250V A F 10A 250V 3 6000 OL 3 x 1 5V AAA 148 mm L 79mm W 48mm H 210g 4 2 18 28 80 600mV 0 1mV 0 5 5 6V 1mV 60V 10mV 600V 1V 10M 600V DC AC 600mV 0 1mV 1 4 6V 1mV 60V 1...

Page 35: ...V 1 2mA 3 2V 30 250V DC AC 9 999Hz 0 001Hz 1 5 5 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999KHz 0 001KHz 99 99KHz 0 01KHz 999 9KHz 0 1KHz 9 999MHz 0 001MHz 200mV 10V AC 250V DC AC 1 99 0 1 2 5 5 200mV 10V AC...

Page 36: ...20 1000 1 1 3 4 1832 1 1 3 250V DC AC 60mA 0 01mA 1 2 5 600mA 0 1mA 10 00A 10mA 3 5 mA F600mA 250V 10A F10A 250V mA 200mA 10A 10A 5A 10 1 60mA 0 01mA 1 5 5 600mA 0 1mA 10 00A 10mA 3 3 mA F600mA 250V 1...

Page 37: ...5 2 F600mA 250V F10A 250V 3 x 1 5V AAA...

Page 38: ...MT 1236 12 0755 83692415 83692986 83246594 83247554 400 169 9629 0755 83692143...

Page 39: ...130 7 5 886 2 22183233 E mail pk mail prokits com tw 1365 25 403 5 0755 83692415 0755 83692143 400 400 1699 629 E mail shenzhen mail prokits com tw...

Page 40: ...PROKIT S INDUSTRIES CO LTD http www prokits com tw E mail pk mail prokits com tw 2020 Copyright by Prokit s Industries Co Ltd C All right reserved...

Reviews: