2.2.4 ACA
RANGE ACCURACY
RESOLUTION
2mA 1uA
20mA
±(1.0%+5)
10uA
200mA ±(2.0%+5)
100uA
20A ±(3.0%+10)
10mA
Max measuring voltage drop
:
200mV.
Max Input current: 20A (less than 10 seconds).
Overload protection
:
0.2A/250V fast-blow fuse, 20A/250V
quick-action fuse for range 20A.
Frequency response: 40Hz-200Hz.
Display: Sine wave RMS (AVG value response).
2.2.5 RESISTANCE
(
Ω
)
RANGE ACCURACY
RESOLUTION
200
Ω
±(0.8%+5) 0.1
Ω
2K
Ω
1
Ω
20k
Ω
10
Ω
200k
Ω
100
Ω
2M
Ω
±(0.8%+3)
1K
Ω
20M
Ω
±(1.0%+15) 10k
Ω
2000M
Ω
±(5.0%(reading-
10)+20)
1M
Ω
Open voltage: less than 3V.
Overload protection: 250V DC or AC peak value.
NOTE:
1. At 200
Ω
range, the test leads should be short-circuited, and
measure the lead to lead resistance, then, subtract from the
real measurement.
2. It is normal to display 10M
Ω
when the test leads shorted in
range 2000M
Ω
, it will not effect the accuracy and shall be
subtracted from the value measured. For example: The object
resistance is 1000M
Ω
, the reading value is 1010M
Ω
, then the
correct value shall be 1010-10=1000M
Ω
3. Lagged display of value is normal when measuring
resistance higher than 1M
Ω
. Please wait until the display is
stable.
3