10. APPENDINX 2. MEASURING PROCESSES
10.1. AC Resistance
AC resistance is measured by applying a small AC signal to TE
module. The AC is generated by a “Commutator” (switch), which
periodically (with 50% duty circle) reverses a circuit of the reference
current
𝐼𝐼
𝑚𝑚
. The “Commutator’s” simplified diagram is shown below.
If there is no input signal, the output voltage of the instrumentation
amplifier equals to
𝐸𝐸
𝑚𝑚
2
⁄
, where
𝐸𝐸
𝑚𝑚
= 4.096 V
.
AC
𝑅𝑅
testing simplified diagram
Output signal of instrumentation amplifier when AC R is tested
During AC resistance measurement the output voltage of the
instrumentation amplifier is sampled and measured by a 12-bit ADC
each time before reversing the current
𝐼𝐼
𝑚𝑚
.
.
The sampling points are
marked as
𝑡𝑡
𝑎𝑎
in the figure above. The voltage drops on TE module for
page 58 of 64