Start the measurement by button <START>.
Result will be indicated with two values:
•
Top value – AC resistance measured as is at ambient
temperature measured by temperature sensor.
•
The bottom value – recalculated AC resistance to standard
reference temperature 27°C (300K). The default reference is
27°C (300K). Optional reference temperature can be set – see
chapter “Main mode”.
Note!
If temperature sensor is not connected to device then
all following measurements to be related to standard
temperature – default value
27°C (300K)
.