Pr
o
Minent
®
30
14
Complies with Directives and Standards
EU directives:
EG-EMV RL 89/336/EWG
91/263/EWG i.d.F. 92/31/EWG
International standards:
EN 50 081-1
EN 50 082-2
15
Technical data
Measured
variable
Chlorine dioxide (ClO
2
)
Application
range
Process water containing surfactants, drinking water
Measurement
range
0.02 - 2 mg/l (at 40
°
C)
pH range
5.5 - 10.5
Temperature
range
10 - 45
°
C, 55 ºC for short periods
Temperature compensated with external temperature sensor Pt 100
Temperature
fluctuation speed
Up to 10 ºC/min
Storage
temperature
5 - 50
°
C
Resolution
Corresponds to measurement range limits below
Max. pressure
3 bar, 1 bar recommended (zero pressure at outlet, no pressure surges)
Sample flow
Recommended:
60 l/h
Minimum:
40 l/h
Maximum:
100 l/h
Cross-sensitivity
Ozone, chlorine
Shelf life
Membrane cap
Typically 3 - 6 months depending upon the water quality
Materials
Membrane cap:
Clear PVC
Electrode shaft:
Black PP and natural coloured PMMA
Supply voltage
16 - 24 V DC
Output signal
4 - 20 mA
Enclosure rating
IP 65
Directives and Standards / Technical data
BA_DT_099_10_08_4spr.p65
13.10.2008, 10:39 Uhr
30