11
Decommissioning
Decommissioning sensor: observe all safety guidelines as given in section
5 “Assembly”.
왘
Disconnect sensor from power supply (see 6 “Installation”).
왘
Depressurise the in-line probe housing assembly.
왘
Loosen the tensioning screw.
왘
Slowly pull out the sensor from the in-line probe housing assembly.
왘
Unscrew the membrane cap over a wash basin or similar and empty.
왘
Rinse away the electrolyte gel with cold water (see 9 “Servicing”).
왘
Rinse the electrodes with clean water and allow to dry.
왘
Screw on a fresh membrane cap loosely to protect the electrodes.
왘
Fit the membrane cap cover to protect the membrane cap.
12
Disposal
Elektrolyte
The electrolyte is non-toxic and can be disposed of with in-house waste.
Sensor
IMPORTANT
•
Electronic waste is classified as special waste.
•
Observe currently applicable local directives.
13
Ordering Information
•
1 bottle of electrolyte (100 ml)
Part no. 1002712
•
1 membrane cap
Part no. 1002710
•
Accessory set
(2 membrane caps + 1 bottle of electrolyte)
Part no. 1002744
•
Two-wire signal cable
(2 x 0.25 mm
2
, Ø 4 mm)
Part no. 725122.6
•
CDP 1-mA-2 ppm (complete set)
Part no. 1002149
The sensors themselves can be ordered only as complete sets:
1 sensor
1 bottle of electrolyte
2 replacement membrane caps
1 operating instructions manual
Decommissioning / Disposal / Ordering Information