Counter output as frequency output
The counter output as frequency output can be used to signal flow or an error, provided it is set
accordingly - see chapter "Settings".
1.
Remove the connected socket.
2.
Remove the insulation of the leads to match the terminals of the other device.
3.
Crimp on a suitable cable end sleeve.
ð
In this format, the cable can be connected to another device.
4.
Connect the cable to both devices in accordance with the wiring diagram. DulcoFlow
®
wiring
diagram - see appendix.
Technical data:
Data
Value Unit
Voltage with open contacts
< 30 V
Current with closed contacts
< 10 mA
Residual voltage with closed contacts
< 0.7 V
Switching frequency
< 10 kHz
Lead
Function
brown
Supply Voltage
white
Error signal
blue
Frequency signal
black
Mass
Assembly and installation
19