CAL pH
Sensor voltage
Sensor calibration in buffer 2
A1018
Buffer temperature
The stability is:
acceptable
good
continue with <CAL>
25.0 °C
173 mV
very good
Fig. 26: Display of the sensor stability achieved
12. The
[acceptable / good / very good] range is displayed
ð
The black part of the horizontal bar indicates the range
detected.
13. Continue with
14.
[Buffer detection] [Manual]: Press and set the buffer value
for buffer 2 to the value of the buffer you are using with the
four arrow keys. Confirm input of the value by pressing
15. Continue with
CAL pH
Asymmetry
Buffer 2:
A1019
Zero point
Slope
% Slope
Accept with <CAL>
Buffer 1:
Calibr.param. for 25 °C
Fig. 27: Display of the calibration result
16.
Incorrect calibration
An error message appears should the result
of the calibration lie outside the specified tol‐
erance limits. In this case the current calibra‐
tion will not be applied.
Check the prerequisites for calibration and
clear the error. Then repeat calibration.
Transfer the result of the calibration into the controller
memory by pressing
ð
The controller displays the continuous display again and
operates with the results of the calibration.
Start up
42