NOTICE!
Seal sensor housing
Possibility of premature drying agent saturation.
A damaged seal can lead to premature saturation of the drying agent.
Check the seal every time the drying agent is exchanged. Correct the seal
seating or exchange it if necessary.
1.
Release the four screws from the corners of the housing and remove the
upper part of the sensor
Fig. 8: Load drying agent
2.
Insert the drying agent immediately after opening the packaging, in
order to avoid premature saturation.
Remove the new drying agent bag (2) from the packaging and place it
together with the humidity indicator card (3) in the lower part of the sensor
(1). In doing so, place the humidity indicator card (3) on the new drying
agent bag (2).