Materials that
come into con‐
tact with the
media
Polyimide (PA), silicone, polypropylene (PP), stainless steel,
borosilicate glass
Voltage supply 100 – 240 VAC, 47 – 63 Hz, 80VA
Galvanic Isola‐
tion
dual insulation, degree of interference overvoltage category
II
Ambient condi‐
tions
Not suitable for operation outdoors.
Operation altitude maximum 2000 m (above sea level).
maximum 95 % relative air humidity (non-condensing).
Enclosure
rating
IP 66
Standards
compliant with
USEPA 180.1 for type
„Infrared“; ISO 7027 and DIN EN
27027 for type
„White light“
Shipping
weight
approx. 2.5 kg
Option: ultrasonic cleaning (TUC 3 /
TUC 4)
The
„Ultrasonic cleaning“ option is
used in order to continuously clean
the sample chamber. It serves neither
to clean previously soiled sample
chambers, nor can it replace manual
cleaning. However, the cleaning inter‐
vals for manual cleaning can be sig‐
nificantly extended by the use of this
system. The use of a sample
chamber with an
„ultrasound
transducer“ is necessary in order to
operate the system. The system emits
an ultrasonic frequency via a spring to
a piezoelectric transducer, which is
connected to the base of the flow-
through cuvette. The cuvette's detec‐
tion system only works in
[AUTO mode].