7.E. Troubleshooting Internal Data Logging
Symptoms
Possible Causes and Comments
The box to the left of the COM port is red.
This means that the COM port is not available.
Another instrument or program could be
occupying the port, making it unavailable. Make
sure to close all programs of this type before
downloading data.
The port selected is incorrect. Click on the box to
the right of the COM port icon to select the
appropriate COM port.
All of the lights are green, but no data are
The connection between the instrument and the
transferred even though the ,
computer is bad. Check and tighten the cable
says, “All data downloaded”
connections. Make sure both ends of the cable
are plugged in tightly.
No data downloaded
IDL not activated. Refer to Section 4 for
instructions.