June 2016
Page 37
8.3
Fuse replacement
WARNING
: Disconnect all the test leads before initiating
the fuse replacement process. Power off the
instrument.
Fuses are located inside the instrument. In order to replace them
you must follow these instructions:
1.
Remove the holster
2.
Unscrew and remove the rear panel using a Phillips suitable
screwdriver.
3.
Remove old fuses and replace them by the new ones.
Fuses must be:
1 A F250V
USING DIFFERENT TYPE OF FUSES COULD
DAMAGE THE INSTRUMENT.
4.
Back to placing the rear panel and screw it.
5.
Put the holster back in place