If the Patch connector cable is disconnected
during a measurement, the monitor shows
screen of Figure 29 . A new measurement can be
started after reconnecting the Patch connector to
the Patch and after pressing the right key on the
monitor (close to the words “New measurement”).
Only in case of pausing a measurement, by
releasing the Recording button, it is possible to
resume the measurement. Then, the EEG data that
is already collected before pausing, is saved.
Step 5
The DeltaScan Monitor will automatically calculate
and present the DeltaScan Score, which is
expressed as:
1, 2, 3, 4 or 5.
(See section 1.5 for an explanation of the scores).
See Figure 30 for an example of a score.
It is strongly advised to write the score down in the
patient record.
Do not rely solely on the DeltaScan Score alone
for starting or adapting delirium treatment (which
may include administration of medication). Clinical
assessment should be leading in deciding on
patient treatment.
Please note!
When the device memory is full, the
oldest fi les are deleted fi rst when starting a new
measurement.
Figure 29: Patch connector got disconnected
Please start a new measurement.
Figure 30: DeltaScan Monitor screen displaying
the result of a measurement.
In relation to defi brillation
Warning!
Remove the DeltaScan Patch and the DeltaScan Monitor from the patient
before applying defi brillation
Warning!
Replace the DeltaScan Patch if it has been on the patient during
defi brillation.