Proline Prosonic Flow 93W
Hauser
7
A0001159
Two-channel measurement: example of arranging sensor pairs at two separate measuring points
A
Measuring channel 1: mounting the sensor pair for measurement via two traverses
B
Measuring channel 2: mounting the sensor pair for measurement via one traverse
Two-path measurement
The flow is measured redundantly at one measuring point in the case of two-path measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-path measurement:
• Individual measured values per measuring channel (output independently of one another)
• The average of the two measured values
The "Averaging" function generally provides you with a more stable measured value. The function is thus
suitable for measurements under conditions that are not ideal (e.g. short inlet runs).
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.
It is generally not necessary to individually configure the two measuring channels in the case of two-path
measurement. However, in certain situations individual channel configuration can be used to balance out
application-specific asymmetries.
A0001160
Two-path measurement: examples of arranging sensor pairs at one measuring point
A
Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via two traverses
B
Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via one traverse
B
A
B
A