Proline Prosonic Flow 93W
12
Hauser
Supply voltage
Transmitter
HART
• 85 to 260 V AC, 45 to 65 Hz
• 20 to 55 V AC, 45 to 65 Hz
• 16 to 62 V DC
PROFIBUS DP/PA
• 9 to 32 V DC
FOUNDATION Fieldbus
• 9 to 32 V DC
Sensor
Powered by the transmitter
Cable entry
Power supply and signal cables (inputs/outputs)
• Cable gland M20 × 1.5
– Cable gland for cables 8 to 12 mm (0.3 to 0.5")
– Cable gland for cables 6 to 12 mm (0.2 to 0.5")
• Thread for cable entry ½" NPT, G ½"
Connecting cable (sensor/transmitter)
Cable gland for one connecting cable per cable entry, 1 ×
∅
8 mm (1 ×
∅
0.31")
• Cable gland M20 × 1.5
• Thread for cable entry ½" NPT, G ½"
Cable gland for two connecting cables per cable entry, 2 ×
∅
4 mm (2 ×
∅
0.16")
• Cable gland M20 × 1.5
• Thread for cable entry ½" NPT, G ½"
A0008152
Cable gland for two connecting cables per cable entry
Connecting cable
(sensor/transmitter)
Only use the connecting cables supplied by Hauser.
Different versions of the connecting cables are available
→
ä
28.
• Cable material made of PVC (standard) or PTFE (for higher temperatures)
• Cable length: 5 to 60 m (16.4 to 196.8 ft)
!
Note!
To ensure correct measuring results, route the connecting cable well clear of electrical machines and switching
elements.
Power consumption
AC: < 18 VA (incl. sensor)
DC: < 10 W (incl. sensor)
Switch-on current
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Lasting min. 1 power cycle
HistoROM/T-DAT save measuring system data if the power supply fails
Potential equalization
For potential equalization, no special measures are necessary.