Proline Prosonic Flow 93W
Hauser
15
Horizontal
In the recommended installation range in a horizontal installation position (View B), gas and air collections at
the pipe cover and problematic deposits at the bottom of the pipe have a smaller influence on measurement.
A0001105
A
Recommended orientation with upward direction of flow
B
Recommended installation range with horizontal orientation
C
Recommended installation range max. 120°
Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the
following inlet and outlet runs is required in order to ensure measuring accuracy.
A0013845
Inlet und outlet runs (top view)
A
Clamp-on version
B
Insertion version
1
= values für single -path version
2
= values für two -path version
1
Valve (2/3 geöffnet)
2
Pump
3
Two pipe bends in different directions
A
B
C
C
³
15 x DN
³
20 x DN
³
20 x DN
³
15 x DN
³
3 x DN
1
2
3
³
15 x DN
³
10 x DN
³
20 x DN
³
20 x DN
³
20 x DN
³
15 x DN
³
15 x DN
³
10 x DN
³
3 x DN
³
3 x DN
1
2
3
A
B
1
1
2
2
1
2
1
2
1
2