Rescue Life User Manual vers. 3.0 Feb. 2015
50
0C to 50C, 15% to 95% non-condensing humidity
Storage Conditions:
-20C to 65C, 15% to 90% non-condensing humidity
Altitude:
Measurement accuracy is not affected by altitude
Auto Interval Periods:
1, 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes
Patient Safety:
Internal operating software ensures that:
Maximum cuff inflation time is limited to 75 seconds
Duration of blood pressure reading is limited to
130 seconds (Adult mode)
120 seconds (Adult Motion Tolerant mode)
90 seconds (Pediatric mode)
75 seconds (Neonate mode)
Additional redundant safety circuitry oversees normal operation and will override to
abort a BP measurement if:
cuff pressure exceeds 300 mmHg (Adult & Pediatric modes) or 150 mmHg
(Neonate mode) at any time
the cuff has been inflated for 180 seconds above 15 mmHg (Adult & Pediatric
modes) or 90 seconds above 5 mmHg (Neonate mode)
Regulatory Standards:
The Module meets all relevant parts of the following Safety/Regulatory Standards:
IEC60601-1
IEC/EN60601-2-30
AAMI SP10
OIML R 16-2
EN1060-1
EN1060-3
EN1060-4