16
FRONT PANEL KEYS
ON/OFF KEY
Power On-Off push button of RESCUE LIFE. At switch on, if the paddles
are disconnected, the battery status and clock set-up screen will appear.
In this case to start ECG monitoring press F1 key.
To access the DATA BASE (only on the AED models) press F3 key(MEM).
To switch off the RESCUE LIFE press once the ON/OFF button. To power-
off the device press the On-Off key only once.
SHOCK KEY
(DEFIBRILLATION)
SHOCK
When the red light inside this key is on it means that RESCUE LIFE is
ready to defibrillate. Pressing this key will release the defibrillation
shock.(This key is active only when disposable pads are used).
To release the shock with the standard pads press both push buttons on
the pads handles.
CHARGE KEY
CHARGE
This key start the charge for the shock.
(This key is active only when disposable pads are used).
To start the charge with the standard pads press both push buttons on
the pads handles.
SPEED DIAL
ENERGY
Control the functional settings of the device.
When pressed, on the screen the parameter to change will be displayed.
By rotating the Speed Dial is possible to change the selected parameter.
Energy is the first parameter that can be selected.
Summary of Contents for Rescue Life
Page 1: ...RESCUE LIFE EXTERNAL BIPHASIC DEFIBRILLATOR AND MONITOR USER MANUAL V 3 0 Feb 2015...
Page 2: ...Rescue Life User Manual vers 3 0 Feb 2015...
Page 6: ...Rescue Life User Manual vers 3 0 Feb 2015...
Page 63: ...Rescue Life User Manual vers 3 0 Feb 2015 57...
Page 74: ......
Page 75: ......
Page 76: ...Rescue Life User Manual vers 3 0 Feb 2015 www progettimedical com...