Rescue Life User Manual vers. 3.0 Feb. 2015
30
W
HEN USING
SYNC
MODE MAKE SURE THAT THE
ECG
TRACE HAS A STABLE BASE LINE AND THE HEART
RATE IS STABLE
. D
EFIBRILLATING IN
SYNC
MODE WITH A DISTURBED
ECG
SIGNAL IS DANGEROUS
BECAUSE THE MACHINE WILL NOT BE ABLE TO IDENTIFY CORRECTLY THE
‘R’
PEAK TO SYNCHRONIZE TO
.
T
HE MACHINE CAN DELIVER A
SYNC
SHOCK USING AS INPUT THE
ECG
CABLE OR THE PADS
,
BUT IT IS
RECOMMENDED TO USE THE PADS INPUT FOR THE BEST RESULT
.