background image

 

 

 

Preva Dental X-Ray System 

 

 
 
 

User Manual 

00-02-1576 Rev. L 

ECN: P1647

 

Summary of Contents for P7015

Page 1: ...Preva Dental X Ray System User Manual 00 02 1576 Rev L ECN P1647...

Page 2: ...s the responsibility of the user to ensure that existing legal regulations regarding installation of the equipment with respect to the building are observed Caution Do not hang lead aprons on the hori...

Page 3: ...More Than Imaging Excellence Progeny Dental 675 Heathrow Drive Lincolnshire Illinois 60069 U S A Phone 888 924 3800 Fax 847 415 9801 WWW progenydental com...

Page 4: ...Preva Dental X Ray System User Manual 00 02 1576 Rev L ECN P1647 Progeny Dental 675 Heathrow Dr Lincolnshire Illinois 60069 U S A Phone 888 924 3800 Fax 847 415 9801 WWW progenydental com...

Page 5: ...Progeny Dental 2006 U S Patents D470237 D469182 D470589 6 837 468 and 6 664 853...

Page 6: ...g an X Ray 9 Using the 12 inch Cone 30 A2033 10 Recommended Maintenance 11 Regular Maintenance 11 Cleaning and Disinfecting 11 Checking System Functions 12 System Function Checklist 13 New Tube Season...

Page 7: ...Articulating Arm allows smooth movement for precise positioning and does not drift or vibrate when left in position Horizontal Arm The Horizontal Arm helps provide the necessary reach for the Preva Th...

Page 8: ...General Information Preva 2 Figure 1 Component Diagram Caution Do not hang lead aprons on the horizontal extension arm...

Page 9: ...n x ray equipment UL 2601 1 File Number E181750 Classified by Underwriters Laboratories Inc with respect to electrical shock fire and mechanical hazards only in accordance with UL 2601 1 and CAN CSA C...

Page 10: ...1 1 IEC 60601 1 2 IEC 60601 1 3 IEC 60601 2 7 IEC 60601 2 28 IEC 60601 2 32 Medical Device directive ISO 13485 EN46001 FDA 510K File Number K043092 Declaration Progeny Inc declares that the products d...

Page 11: ...l Electrical code or VDE standards concerning provisions of an additional protective earth ground terminal for power supply connection Before cleaning or disinfecting this equipment must always be dis...

Page 12: ...DU PERSONNEL AUTORISE VOIR MANUEL DE L OPERATEUR WARNING X RAY THIS X RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED X RAY EMISS...

Page 13: ...e Preva Dental X Ray System Using the Operator Panel Power On Settings When the Preva Dental X Ray System is powered on the Operator Panel selections are those that were in use when the system was las...

Page 14: ...sed and held until the Radiation Indicator no longer illuminates and the audible signal is no longer heard Releasing the Exposure button immediately terminates the x ray exposure CAUTION Releasing the...

Page 15: ...down arrow buttons to adjust the value Skip this step if you are using pre programmed exposure settings Note When exposure settings are being adjusted the Tooth Selection Image Receptor Type and Patie...

Page 16: ...y System is factory set for use with the standard supplied 8 inch 20 cm Cone The 12 inch 30 cm Cone 30 A2033 is available Using the longer cone requires longer exposure times See the System Configurat...

Page 17: ...Protect the system from contamination using barriers available from dental distributors Follow the disinfectant manufacturer s recommendations when using their cleaner or disinfectant Cleaning Method...

Page 18: ...enance as indicated in the User Manual Failure to perform these checks may result in an installation that does not comply with U S Radiation Performance Standards 21 CFR Subchapter J CAUTION If the Pr...

Page 19: ...Controls With the power switch located at the upper right of the Control Unit in the ON position verify that technique factors appear on the Operator Panel Also check the function of the selection but...

Page 20: ...turning to normal operation any time the system has been unused for more than two months 1 Verify system operation 2 Energize the system 3 Select 60 kilovolts 7 milliamperes and the exposure time of o...

Page 21: ...posure time 3 Inspect the condition of the remaining imaging chain components such as the film chemistry and processor or the condition of the x ray sensor and computer No X Ray If no x ray is produce...

Page 22: ...7 7 7 7 7 7 7 7 Lower Molar Seconds 0 125 0 064 0 100 0 050 0 160 0 080 0 125 0 080 0 400 0 200 0 200 0 100 KV 65 65 60 60 60 60 60 60 60 60 60 60 MA 7 7 7 7 7 7 7 7 7 7 7 7 Upper Molar Seconds 0 160...

Page 23: ...own arrow highlight Configure Unit and press the right arrow key See Figure 34 3 Select Languages on the next screen is highlighted press the right arrow key 4 Select the desired language and press th...

Page 24: ...nter system configuration mode depress the Tooth Selection and Patient Size Selection buttons on the Operator Panel simultaneously for 5 seconds The display shows the Main System Configuration menu as...

Page 25: ...splay to the Main System Configuration menu shown in Figure 3 Adjusting Contrast 1 Select ADJUST CONTRAST from the menu You will see the Progeny logo 2 Use the up and down arrows to increase or decrea...

Page 26: ...changing presets use the table below to write down the presets you are programming Displaying the Preset Options Menu 1 From the Main System Configuration menu shown in Figure 3 select CHANGE PRESETS...

Page 27: ...g Edit Preset Mode and Tooth Size Image Receptor Type and Patient Size are illuminated 2 Use the Tooth Selection Image Receptor Type and Patient Size Selection buttons to select the preset to change T...

Page 28: ...MA 6 6 7 7 7 7 7 7 7 7 7 7 Lower Molar Seconds 0 080 0 040 0 080 0 040 0 160 0 080 0 100 0 050 0 080 0 040 0 250 0 125 KV 70 70 65 65 60 60 60 60 65 65 65 65 MA 6 6 7 7 7 7 7 7 7 7 7 7 Upper Molar Se...

Page 29: ...desired adjust the settings using System Configuration mode Record your settings in the table below 8 inch 20 cm Cone Digital Receptor D speed Film E F Speed Film Tooth Selection Setting Adult Child A...

Page 30: ...20 cm Cone Digital Receptor D speed Film E F Speed Film Tooth Selection Setting Adult Child Adult Child Adult Child kV mA Incisor seconds kV mA Bicuspid seconds kV mA Bitewing seconds kV mA Lower Mol...

Page 31: ...rmational only 1 From the Main System Configuration menu shown in Figure 3 select CONFIGURE UNIT You will see the Configuration menu shown in Figure 6 2 Select SHOW CONFIG The display will show Curren...

Page 32: ...Cone size in the Set Configuration menu Using a 12 inch Cone 1 From the Main System Configuration menu shown in Figure 3 select CONFIGURE UNIT You will see the Configuration menu shown in Figure 6 2 S...

Page 33: ...s for that exposure Until you exit diagnostic mode the display will continue to show feedback values after each exposure 1 From the Main System Configuration menu shown in Figure 3 select CONFIGURE UN...

Page 34: ...m 12 inch 30 cm Minimum Half Value Layer 1 7 mm Aluminum equivalent at 70 kVp Minimum Inherent Filtration 2 mm Aluminum equivalent 70 kVp Focal Spot 0 4 mm IEC 336 Automatic Cooling Time 15 times the...

Page 35: ...Specifications Preva 29 Figure 8 Tube Rating Charts...

Reviews: