background image

Operating Instructions                                                                                                                                    JB-70 
 

 
 

 
 
 

10 

Ready Indicator 

The Ready Indicator illuminates when the system is ready to make an exposure. 
Immediately after an exposure, the Ready Indicator flashes until the x-ray tube 
cools down sufficiently to make the next exposure. When the Ready Indicator is 
flashing, no exposure can be made. 

Radiation 
Indicators 

The JB-70 has a visible and an audible Radiation Indicator. When an exposure is 
in progress, the Radiation Indicator on the Operator Panel is illuminated and a 
audible tone is heard. The exposure is complete when the Radiation Indicator is 
extinguished and the audible tone is no longer heard. 

 

Taking an X-Ray 

 

1. 

Turn the power switch, located on the base of the Control Unit, to the “On” 
position. The Ready Indicator on the front of the Operator Panel will light. 

2.  Exposure time can be selected by either using the pre-programmed 

exposure times or by simply adjusting the time manually up or down to the 
desired time. If selecting the time manually, skip steps 3-5. 

3.  Verify that the unit is set for the correct Image Receptor Type. The icon for 

the currently selected Image Receptor Type is illuminated. To change the 
Image Receptor type, press the Image Receptor Type button until the correct 
Image Receptor Type is selected. 

4.  Verify that the system is set for the appropriate Patient Size. The icon for the 

currently selected Patient Size is illuminated. The change the Patient Size, 
press the Patient Size button until the correct Patient Size is selected. 

5.  Verify that the unit is set for the Tooth to be imaged. The icon for the 

currently selected Tooth is illuminated. To change the Tooth Selection, press 
the Tooth Selection button until the correct Tooth is selected.  

6.  If desired, the Exposure Time Selection up or down buttons can be used to 

change the pre-set exposure time. 

Note:

 When the Time Selection buttons are used, the Tooth Selection, 

Image Receptor Type and Patient Size buttons are turned off. 

7.  Position the Tubehead to the patient using standard accepted positioning 

procedures.  

8.  Press and hold the Exposure button until the audible signal is no longer 

heard and the Radiation Indicator is no longer illuminated. Releasing the 
Exposure button or coil-cord hand switch at any time will immediately 
terminate the exposure. 

Note: 

When using the coil-cord hand switch, it is recommended that the 

operator exit the operatory if possible. 

Note:

 In order to comply with regulations and good safety practices, the 

technique factors must be visible to the operator from the remote location.

 

9.  Return the Tubehead to the storage position. 

Note:

 Be careful not to bang the Tubehead on the wall when returning it to 

the storage position. 

 

Summary of Contents for I7016

Page 1: ...JB 70 Dental X Ray System User Manual 00 02 1568 Rev A Progeny Dental 1407 Barclay Boulevard Buffalo Grove Illinois 60089 U S A Phone 888 924 3800 Fax 847 459 5175 WWW progenydental com...

Page 2: ...Progeny Dental 2002 Patents Pending...

Page 3: ...JB 70 Dental X Ray System 9 Operator Panel Controls 9 Taking an X Ray 10 Using the 12 in Cone 30 A2033 11 Recommended Maintenance 12 Regular Maintenance 12 Cleaning and Disinfecting 12 Checking Syste...

Page 4: ...Table of Contents JB 70 ii...

Page 5: ...e Articulating Arm provides the articulation support for the Tubehead and the reach and coverage of the Tubehead to the patient The Articulating Arm allows smooth movement for precise positioning and...

Page 6: ...General Information JB 70 4 Figure 1 Component Diagram...

Page 7: ...ment UL 2601 1 File Number E181750 Classified by Underwriters Laboratories Inc with respect to electrical shock fire and mechanical hazards only in accordance with UL 2601 1 and CAN CSA C22 2 NO 601 1...

Page 8: ...ollowing regulatory documents apply 21CFR UL 2601 1 IEC 60601 1 2 IEC 60601 1 3 IEC 60601 2 7 IEC 60601 2 28 IEC 60601 2 32 Medical Device directive ISO 13485 EN46001 Declaration Progeny Inc declares...

Page 9: ...tional Electrical code or VDE standards concerning provisions of an additional protective earth ground terminal for power supply connection Before cleaning or disinfecting this equipment must always b...

Page 10: ...ATION SEULEMENT PAR DU PERSONNEL AUTORISE VOIR MANUEL DE L OPERATEUR WARNING X RAY THIS X RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE...

Page 11: ...can also be adjusted using the up and down Exposure Time arrows For a table of the pre programmed exposure times set at the factory refer to the Pre programmed Exposure Times table later in this manua...

Page 12: ...eceptor Type is selected 4 Verify that the system is set for the appropriate Patient Size The icon for the currently selected Patient Size is illuminated The change the Patient Size press the Patient...

Page 13: ...one 30 A2033 is recommended when using the paralleling film positioning technique Using the long cone requires the use of longer exposure times which can be programmed into the system by changing an i...

Page 14: ...roblems are resolved Regular Maintenance In the interest of equipment safety a regular maintenance program must be established This maintenance program should consist of cleaning and disinfecting as w...

Page 15: ...is in the ON position Operator Panel Controls With the power switch located at the base of the Control Unit in the ON position verify that a time indicates on the Operator Panel Also check the functi...

Page 16: ...programmed exposure times make sure that the proper image receptor type is selected 2 Verify proper operation of the film processor No X Ray If no x ray is produced check the following 1 Verify that t...

Page 17: ...an be modified using the procedure that follows the table Times for 8 inch 20 cm Cone in Seconds 60 Hz 50 Hz times shown in parentheses Digital Receptor D speed Film E F Speed Film Tooth Selection Adu...

Page 18: ...ld Incisor 0 133 0 160 0 100 0 120 0 417 0 500 0 200 0 240 0 233 0 280 0 117 0 140 Bicuspid 0 167 0 200 0 083 0 100 0 533 0 640 0 267 0 320 0 333 0 400 0 167 0 200 Bitewing 0 200 0 240 0 100 0 120 0 4...

Page 19: ...e time is to be changed a Use the Image Receptor Type button to select the receptor type b Use the Tooth Selection button to select a tooth c Use the Patient Size button to select the patient size 3 U...

Page 20: ...50 Hz operation Timer Accuracy 34 ms Source to Skin Distance 8 inch 20 cm 12 inch 30 cm Radiation Output Approximately 800 mR at the following sets 70 kVp 8 mA 1 0 sec 20 cm SID Minimum Inherent Filtr...

Page 21: ...adiation produced during this time is small it is useful and contributes to forming the image on the film When the exposure time starts the first several x ray pulses will be high kVp about 80 low mA...

Reviews: