3302C Series Mainframe Operation Manual 4-15
4.waveform bank 10 is DC.
5.There are five (5) waveform information for each waveform bank, therefore, there
is a total 55 waveform information for eleven (11) waveform banks. Waveform
information is shown in Table 4-4.
For Example
:
a.BANK 1, to set waveform bank 1.
b.BANK 10, to set waveform bank 10.
WAVE FROM
BANK
A
B
C
D
E
0
√
2 2.0 2.5 3.0 3.5
1
1.5 1.6 1.7 1.8 1.9
SINE WAVE
2
3.0 3.1 3.2 3.3 3.4
C.F.= 2.0
3
P.F.=-0.85 P.F.=-0.80 P.F.=-0.75 P.F.=-0.70 P.F.=-0.65
C.F.= 2.5
4
P.F.=-0.70 P.F.=-0.65 P.F.=-0.60 P.F.=-0.50 P.F.=-0.40
C.F.= 3.5
5
P.F.=-0.50 P.F.=-0.45 P.F.=-0.40 P.F.=-0.35 P.F.=-0.30
C.F.= 2.0
6
P.F.= 0.85 P.F.= 0.80 P.F.= 0.75 P.F.= 0.70 P.F.= 0.65
C.F.= 2.5
7
P.F.= 0.70 P.F.= 0.65 P.F.= 0.60 P.F.= 0.50 P.F.= 0.40
C.F.= 3.5
8
P.F.= 0.50 P.F.= 0.45 P.F.= 0.40 P.F.= 0.35 P.F.= 0.30
SQUARE
WAVE
9
1.0 1.1 1.2 1.3 1.4
DC
10
√
2dc 2dc 2.5dc 3.0dc 3.5dc
Table 4-5 waveform information
WAVEFORM
Purpose
:
To set waveform, for 3250A series electronic load module.
Command syntax
:
WAVE{SP}{m}{
;
|NL}m
:
0
~
4
Description
:
This command is to set the current C.F. at CC MODE (Peak Value Factor).
This command works only at CC MODE. When BANK varies, these 5 sets of C.F.
will at the same time define different C.F. as shown in Table 4-4. For details,
please refer to 3250 Operation Manual.
For Example
:
a.WAVE 1, to set 2nd set C.F.
b.WAVE 4, to set 5th set C.F.
FREQUENCY
Purpose
:
Setting of Frequency Value for 3250 series electronic load module.
Command syntax
:
FREQ{SP}{NR2}{
;
|NL}