4-12 PRODIGIT
LEVEL HIGH/LOW
Purpose
:
Select Low or High level in static mode of 3310 series electronic load, or LEVEL
A/B of 3250A series electronic load.
Command syntax
:
LEVE {SP}{HIGH|LOW|A|B}{NL}
Description
:
LEVE LOW is Set LOW current level in CC mode, LOW resistance level in CR
mode, or LOW voltage level in CV mode at the active load channel.
LEVE 1 is Set HIGH current level in CC mode, HIGH resistance level in CR mode,
or HIGH voltage level in CV mode at the active load channel.
LEVE B is The condition and setting of 3250A series electronic load adopt the set
value of LEVEL B.
PRESET ON/OFF
Purpose
:
Set the upper or lower 4 1/2 digit multi-function meter to display the programming
load level.
Command syntax
:
PRES{SP}{0|1|OFF|ON}{NL}
Description
:
PRES ON is set the load module in the 3302C mainframe to preset on status.
MODE CC/CR/CV/LIN/CP
Purpose
:
Select the operating mode of Electronic load module.
Command syntax
:
MODE{SP}{0|1|2|3|CC|CR|CV/LIN|CP}{NL}
Description
:
MODE CC is set the presently operating mode to Constant Current mode for the
load module in the 3302C mainframe.
MODE CV is set the presently operating mode to Constant Voltage mode.
MODE 1 is set the presently operating mode to Constant Resistance mode.
MODE CP is set the presently operation mode to Constant Power mode.
Note
:
MODE CV is available in 3310A series only.
MODE CP is available in 3310C series only.
CLEAR status register
Purpose
:
CLEar the PROT and ERR status byte registers.
Command syntax
:
CLR{NL}
Description
: