![Proceq Resipod Family Operating Instructions Manual Download Page 8](http://html1.mh-extra.com/html/proceq/resipod-family/resipod-family_operating-instructions-manual_1608684008.webp)
8
© 2017 Proceq SA
In order to enable this, the Resipod is also supplied with large surface area foam contact pads .
Simply snap out the steel contacts and replace with the foam pads . They should likewise be moist-
ened before measurements .
Hold and Save function
Once a stable reading has been achieved, click the hold button on the side of the Resipod to freeze
the actual measurement on the screen .
Hold
• The display flashes to indicate the “hold” status .
• Click the hold button again to return to “live” mode or:
Save
• Click the save button to record the measurement . An “m” appears to indicate that a new
reading has been stored in this memory location . (In the example, the reading is the first reading
stored in memory object 2 .)
3.2 Memory function
The Resipod can store up to 512 measurements . The memory is arranged so that the readings are
stored in objects from 1 to 19 . Each object can contain up to 99 readings .
Object 1 - Reading 1
Object 19 – Reading 99
You may check the last entry “Object – Reading” number simply by pressing the “Save” button
when the instrument is not in hold modus .