Proceq Resipod Family Operating Instructions Manual Download Page 17

17

© 2017 Proceq SA

6 Units, Parts and Accessories

Part No.

Description

381 10 000

Resipod, 50mm probe spacing, test strip, foam contact pads, charger with USB-

cable, software, carrying strap, documentation and case .

381 20 000

Resipod, 38mm (1 .5”) probe spacing, test strip, foam contact pads, charger with 

USB-cable, software, carrying strap, documentation and case .

381 30 000

Resipod Bulk Resistivity, 50mm probe spacing, test strip, foam contact pads, 

charger with USB-cable, software, carrying strap, documentation and case, 

Bulk Resistivity Accessory .

381 40 000

Resipod Bulk Resistivity, 38mm (1 .5”) probe spacing, test strip, foam contact 

pads, charger with USB-cable, software, carrying strap, documentation and 

case, Bulk Resistivity Accessory .

381 50 000

Resipod Geometric, 50mm probe spacing, test strip, foam contact pads, char-

ger with USB-cable, software, carrying strap, documentation and case, Resipod 

Geometric Accessory .

381 60 000

Resipod Geometric, 38mm probe spacing, test strip, foam contact pads, 
charger with USB-cable, software, carrying strap, documentation and case, 
Resipod Geometric Accessory .

Parts and Accessories

Description

381 01 088

Bulk Resistivity Accessory

381 01 098

Resipod Geometric Accessory

381 01 094

Variable Spacing Probe

381 01 089

Resipod Stand

381 01 043 S

Set of replacement foam contact pads (5x4)

381 01 092 S

Bulk resistivity contact pads . Set of 10 pieces

381 01 038

Test strip

381 01 031

Inner contact cap (See note below)

381 01 041

Outer contact cap (See note below)

381 01 033

Contact complete (See note below)

381 01 036

Contact spring

381 01 014

USB cover

381 01 070

Key for USB port

391 80 110

Carrying strap

 

To replace one of the inner contacts requires parts 381 01 031, 381 01 033 and 381 01 036 .
To replace one of the outer contacts requires parts 381 01 041, 381 01 033 and 381 01 036 .

7 Technical Specifications

Measurement range:

1k

Ω

cm - ca . 1000 k

Ω

cm (dependent on probe spacing) 

Resistance measurement accuracy:

Nominal current 200µA

±0 .2 k

Ω

cm or ±1% (whichever is greater)

Nominal current 50µA

±0 .3 k

Ω

cm or ±2% (whichever is greater)

Nominal current <50µA

±2 k

Ω

cm or ±5% (whichever is greater)

Frequency

40 Hz

Memory

Non volatile, ca . 500 measured values

Power Supply

>50 hours autonomy

Charger connection

USB type B, (5V, 100mA)

Dimensions

197 x 53 x 69 .7 mm (7 .8 x 2 .1 x 2 .7 inch)

Weight

318 g (11 .2 oz)

Operating temperature

0° to 50°C (32° to 122°F)

Storage temperature

-10° to 70°C (14° to 158°F)

Summary of Contents for Resipod Family

Page 1: ...Made in Switzerland Operating Instructions Concrete Durability Testing family...

Page 2: ...ring Resistivity with Resipod 7 3 1 Making a Measurement 7 3 2 Memory Function 8 4 Resistivity Measurements on Site 10 4 1 Influences 10 4 2 Applications 11 5 Quality Control Applications 13 5 1 Resip...

Page 3: ...resented in good faith and believed to be correct Proceq SA makes no warranties and excludes all liability as to the completeness and or accuracy of the information Safety Instructions The instrument...

Page 4: ...Resipod Measurement Principle The Resipod is an evolution of the industry standard CNS Farnell RM MKII resistivity meter operating on the principle of the Wenner probe The Resipod is designed to measu...

Page 5: ...ffset against the need to avoid the influence of reinforcing steel see chapter 3 50 mm spacing is typically considered a good compromise Both units operate with a digitally generated 40 Hz alternating...

Page 6: ...ed in outer probes and the reading is rounded to the nearest k cm This mode works down to a current of 10 A 1 segment lit Below this range OL is indicated See Indication of poor connection below Curre...

Page 7: ...its bottom this will fill the reservoirs Press the Resipod firmly down until the outer two rubber caps rest on the surface to be tested Indication of poor connection In the case of a poor connection...

Page 8: ...e hold status Click the hold button again to return to live mode or Save Click the save button to record the measurement An m appears to indicate that a new reading has been stored in this memory loca...

Page 9: ...the air for this the open line screen will be displayed and will blink Press the Save button for 2 seconds to delete the reading Object 4 Reading 3 has been deleted The display shows a small c to ind...

Page 10: ...to the rebars as shown RILEM TC154 EMC ELECTROCHEMICAL TECH NIQUES FOR MEASURING METALLIC CORROSION recommends making 5 readings from the same location moving the probe a few mm between each measure m...

Page 11: ...pirical tests have arrived at the following typical values for the measured resistivity which can be used to determine the likelihood of corrosion These figures are for Ordinary Portland Cement at 20...

Page 12: ...concrete structure will identify the most permeable areas Such areas are more likely to be susceptible to chloride penetration The Resipod model with 38 mm spacing conforms to the AASHTO T 358 Surface...

Page 13: ...m with a maximum aggre gate size 1 5 38mm Fixed probe spacing 1 5 38mm Bulk resistivity test on cylinders up to 100mm 4 diameter Surface resistivity test on non standard cylinders with ag gregate size...

Page 14: ...cylinder but they also have a resistance which must be measured and compensated for to determine the true bulk resistivity of the cylinder under test The resistance of the foam inserts varies with the...

Page 15: ...te Lower foam insert Lower plate Test set up to measure the bulk resistivity of the cylinder Rmeasured Upper plate Upper foam insert Cylinder under test Lower foam insert Lower plate The measured resi...

Page 16: ...ipod instrument so that the display shows directly the correct value in k cm See Chapter 9 5 2 Resipod Geometric Resipod Geometric is supplied with a stand for mounting the Resipod and a variable spac...

Page 17: ...sories Description 381 01 088 Bulk Resistivity Accessory 381 01 098 Resipod Geometric Accessory 381 01 094 Variable Spacing Probe 381 01 089 Resipod Stand 381 01 043 S Set of replacement foam contact...

Page 18: ...l check using the full 200 A range The lower row allows the functional check using the reduced 50 A range Resistors to the outer pads limit the maximum current to approx 20 A only two segments in the...

Page 19: ...dirt or salt deposits might create leakage currents and yield faulty readings Note The springs are intentionally long to ensure a con stant pressure on all four probes while allowing a 4 mm displacem...

Page 20: ...the start menu The Resipodlink starts with a blank list Application settings The menu item File Application settings allows the user to select the language and the date and time format to be used Conn...

Page 21: ...series from the summary click on the summary symbol in the impact counter column This symbol is either black or greyed out which shows whether or not the series is included in the summary Adjusting t...

Page 22: ...sipod only data in the current proj ect Restoring original downloaded data Select the menu item File Restore all original data to restore the data to the original format as it was downloaded This is a...

Page 23: ...enu item Device Set measurement settings allows a correction factor to be downloaded onto the instrument so that it gives a direct readout of the resistivity when being used with The Variable Spacing...

Page 24: ...inders k 2 1 09 0 527 7 34 d a d a d diameter of the cylinder mm a probe spacing mm L length of the cylinder mm Formula valid for d a 4 and L a 5 e g for a 200mm cylinder the maximum probe spacing all...

Page 25: ...icate that a correction factor has been set Applied correction 2 a k a in cm Display of corrected resistivity values in ResipodLink The correction settings are recorded with the measurement data and d...

Page 26: ...26 2017 Proceq SA Notes...

Page 27: ...27 2017 Proceq SA Notes...

Page 28: ...7 info asia proceq com Proceq Rus LLC Ul Optikov 4 korp 2 lit A Office 410 197374 St Petersburg Russia Phone Fax 7 812 448 35 00 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharja...

Reviews: