6
© 2015 Proceq SA
2. Technical Specifications
Instrument
Range
0.1 – 7930 μs
Resolution
0.1 μs (< 793 μs), 1 μs (> 793 μs)
Display
7” colour display 800x480 pixels
Pulse Voltage UPV
100 Vpp – 450 Vpp
Pulse Voltage UPE
100 Vpp - 400 Vpp
Receiver Gain
1x – 10’000x (0 – 80dB) [11 steps]
Receiver Sensitivity
10 μV
Receiver Input Impedance
7 k
Ω
Pulse Echo Range
0.1 – 1200 μs
Transducer Frequency
50 kHz
Aperture Size
2x25 cm
2
Bandwidth
20 – 500 kHz
Memory
Internal 8 GB Flash memory
Regional Settings
Metric and imperial units and
multi-language supported
Battery
Lithium Polymer, 3.6 V, 14.0 Ah
Mains
9 V – 15 V / 2.0 A
Weight
About 1525 g (incl. Battery)
Dimensions
250 x 162 x 62 mm
Operating Temperature
0°C – 30°C (Charging*, running instrument)
0°C – 40°C (Charging*, instrument is off)
-10°C – 50°C (Non-charging)
Humidity
< 95 % RH, non condensing
IP Classification
IP54
Standards and Directives
CE certification
Battery Lifetime
> 8h (in standard operating mode)
Pollution Degree
2
Installation Category
2
*charging equipment is for indoor use only (no IP classification)
Power Supply
Model
HK-AH-120A500-DH
Input
100-240 V / 1.6 A / 50/60 Hz
Output
12 V DC / 5 A
Max. Altitude
2’500 m above sea level
Humidity
< 95%
Operating Temperature
0°C - 40°C
Environment
Indoor use only
Pollution Degree
2
Installation Category
2