9
© 2009 by Proceq SA
Display warnings
3.7
Review list
A previous series can be reviewed directly from the data list. The lower portion of the data list shows
the last 20 completed measurement series identified by name.
This symbol identifies a file which contains data.
Scroll to the series you wish to review and press the select button. You are given the option to look at
the summary data or to delete the series. If you choose to delete the series, the confirmation window
will be displayed.
If you wish to look at the summary data, you will be taken
directly to the statistics screen. (See 3.4)
Note: Only the last 20 completed measurement series can be reviewed in the data
list. The PaperSchmidt stores many more. The remaining series can only be seen
once the data has been put into Paperlink.
4. Technical Specifications
Mechanical Data
Impact energy
0.735 Nm
Hammer mass
115 g
Spring constant
0.262 N/mm
Spring extension
75 mm
Dimensions of housing
55 x 55 x 250 mm (340 mm to tip of plunger)
Dimensions (plunger visible part)
105 x ø15 mm (radius of spherical tip 25mm)
Weight
570 g
Memory Data
No. of impacts in a series
Maximum 99
No. of series names
50 series names. Each name can be up to 12 characters long.
Memory capacity
Dependent on length of test series, Example 1: 401 series with 10
values per series, Example 2: 246 series with 20 values per series
Electrical Data
Display
17 x 71 pixel, graphic
Power consumption
~13 mA measuring, ~4 mA set-up and review, ~0.02 mA idle
Accumulator duty
> 5000 impacts between charges
Charger connection
USB type B (5V, 100 mA)
Accumulator capacity
~150 mAh
Environmental conditions
Operating temperature
0 to 50°C
Storage temperature
-10 to 70°C
When carrying out a measurement (See 3.3) the display
will give a warning if any of the limits are exceeded. In
this example, the limit sign next to the R indicates that the
lower limit has been exceeded. If the upper limit has been
exceeded, the arrow points upwards.
In the roll profile view (after TIP function and in the sum-