CHANGE TRANSDUCER TIP
1. Remove adjusting sleeve
1
and
locking ring
2
2. Unscrew the brass nut
3
3. Pull out the spring
4
4. Pull out the complete tip unit
5. Remove the teflon
cover
5
if mounted
6. Unscrew the tip
6
from the core
7
and mount the new transducer tip
7. Remount in opposite order
2
3
5
6
7
1
4
MAINTENANCE
• The calibration certificate supplied with the Prisma DI-5 is valid
for 3 years. Please contact us if you would like to re-calibrate it.
• Make sure that the extension bars and measuring tips are in good
condition and replace it if it’s needed.