280490 FGSH-450
- Owner’s Manual
Rev: 03/25/13
Page: 22
General Care, Inspection and Maintenance… continued
This section to be only completed by a qualified service technician as authorized by Prism
Medical Ltd. Use the provided service record history sheets to note findings during inspection.
A) Each Use - To be completed by User
Hydraulic Pump, and Leg Spreading are operational.
B) Semi-Annually - To be completed at least every 6 months.
Should any of the items as per check list on next page fail the inspection do not use the lift. Contact
Prism Medical Systems or your local qualified service technician for service.
Complete the inspection as per check list on next page.
Check and lubrication at every six months to all pivot points as listed on previous page.
C) Annually - To be completed at least every 12 months.
Complete the inspection as noted in the “Semi-Annually” section above.
Also check the following:
A light grease (water proof auto lubricant) may be applied to the ball bearing swivel of the
casters once a year.
Perform a working load test of one (1) lifting cycle with the maximum load. This is in accord-
ance to the periodic inspection requirements of ISO10535.