CARRY BAR ATTACHMENT WITH MOUNTING BRACKET:
1. Insert the carry bar pin and washers up through the swivel bar pad.
2. Align the hole of carry bar pin with the hole of mounting bracket and secure with shoulder bolt and nut
provided as shown in Figure 14.
Figure 14
Figure 15
Figure 16
NOTE: Make sure that carry bar pin and washers are completely pushed up through the Carry
Bar Pad.
The carry bar has hooks at both ends to attach the sling straps that supports the patient
during the lift. The carry bar pad should remain in place during normal use.
After assembly, check to ensure that:
1. The mast is fully locked into the tube of base assembly.
2. Hydraulic pump/actuator is moving up and down using the pump handle.
3. Control valve is working properly.
4. Leg spreading is working properly using the shifter handle.
5. Rear wheel brakes works properly.