Technical manual August 2009
Rev: 01/08/09
Page:
30
LCD Display & Programming Functionality
−
If the hoist has the return-to-charge (“RTC”) feature, the following programming choices
will also be available:
1.– RTC Max. Time
(i)
60 seconds
(ii)
120 seconds (factory setting)
(iii)
180 seconds
(iv)
240 seconds
2.– RTC Drop Time
(i)
9 seconds
(ii)
12 seconds
(iii)
15 seconds
(iv)
18 seconds
(v)
21 seconds
(vi)
24 seconds
3.– RTC Speed
(i)
2
(ii)
4
(factory setting)
(iii)
8
•
Measuring Lifts
−
The hoist should add one to the lift counter if the hoist has been operated under load
(i.e., approx. 60 lbs. with 2’ of travel).
•
Preventative Maintenance
−
Preventative maintenance should be completed every six (6) months. The hoist should
recommend preventative maintenance if it hasn’t had any preventative maintenance for:
1.– 1,000 lifts (four or five lifts a day 180 days); or,
2.– Five (5) hours.
−
To recommend preventative maintenance, the hoist will:
17.
Beep one (1) time every thirty (30) minutes, this can be silenced or over-ridden
by the user by changing a setting, see above.
18.
Flash “Maintenance” in the first line of the display (regardless of which default
display mode the user has selected).
M a i
t e n
n
a n c e
Flashing
−
To reset the counter which notifies the hoist when to signal for preventative
maintenance, see the following section titled, “Resting Lift Counter”.