Chapter 6
Advanced Topics
61
Figure 50. Two Examples illustrating Partial Illumination of CCD for Kinetics Mode
Note:
The illuminated area is the furthest from the serial register.
•
The left image shows an illuminated area of 60 rows. Window Size is 60 rows high.
•
The right image shows an illuminated area of 12 rows. Window Size is 12 rows high.
ß
à
t
exp
#1
ß
à
t
s
#1
ß
à
t
exp
#2
ß
à
t
exp
#n
ß
à
t
s
#2
ß
à
t
s
#n
Exposure/
Subframe
Vertical Shift
Readout
~500 msec
@
1 MHz readout
~66 msec
@
10 MHz readout
(For a 512 x 1048 frame)
Figure 51. Kinetics Data acquired based on Masked Images shown in Figure 50.
(For more information, see Notes on next page.)
Summary of Contents for PhotonMAX System
Page 1: ...4411 0107 Version 1 B January 26 2006 4411 0107...
Page 12: ...12 PhotonMAX System Manual Version 1 B This page intentionally left blank...
Page 20: ...20 PhotonMAX System Manual Version 1 B This page intentionally left blank...
Page 66: ...66 PhotonMAX System Manual Version 1 B This page intentionally left blank...
Page 78: ...78 PhotonMAX System Manual Version 1 B This page intentionally left blank...
Page 81: ...This page intentionally left blank...
Page 82: ...82 PhotonMAX System Manual Version 1 B This page intentionally left blank...
Page 88: ...88 PhotonMAX System Manual Version 1 B PhotonMAX Camera C mount Liquid Cooled...
Page 89: ...Appendix B Outline Drawings 89 CoolCUBE Circulator...
Page 90: ...90 PhotonMAX System Manual Version 1 B This page intentionally left blank...
Page 98: ...98 PhotonMAX System Manual Version 1 B This page intentionally left blank...