5
PART NUMBER
DESCRIPTION
DATE INSTALLED
1050714
PSLT 16.7ftWC with 50ft cable
1050713
PSLT 16.7ftWC with 75ft cable
1050712
PSLT 16.7ftWC with 100ft cable
1050717
PSLT 33.4ftWC with 50ft cable
1050716
PSLT 33.4ftWC with 75ft cable
1050715
PSLT 33.4ftWC with 100ft cable
*Backed by a 2-year limited warranty.
PSLT Analog Input Scaling Table
mA
%
TRANSMITTER 0-16.7FTWC
TRANSMITTER 0-33.4FTWC
Ft (cm)
In (mm)
Ft (cm)
In (mm)
20
100
16.7 (509)
200 (5,090)
33.4 (1,180)
400 (10,180)
16
75
12.5 (382)
150 (3,820)
25 (764)
300 (7,635)
12
50
8.4 (254)
100 (2,540)
17 (508)
200 (5,090)
8
25
4.2 (127)
50 (1,270)
8.5 (254)
100 (2,540)
4
0
0 (0)
0 (0)
0 (0)
0 (0)
844-4PRIMEX (477-4639)
WWW.PRIMEXCONTROLS.COM
© 2016 SJE-Rhombus
®
PN 1051390B
PRIMEX
®
is a trademark of SJE-Rhombus
®
Viton® is a trademark of E. I. DuPont de Nemours and Company Corporation
SPECIFICATIONS
•
Measuring Ranges: 7.25 PSI (16.7 ftWC) and 14.50 PSI (33.4 ftWC)
•
Accuracy: 1% full scale
•
Operating Temp: 32-122°F (0-50°C)
•
Output: 4-20 mA, loop powered
•
Excitation Voltage: 8-32 VDC reverse polarity protection
•
Overpressure Protection: 58 PSI (133.6 ftWC)
•
IP Rating: IP 68 for permanent submersion
•
Wetted Materials:
•
Sensor Housing: PVC, 2.00” diameter x 5.62” length (51mm x 143mm)
•
Cable Seal: EPDM
•
Diaphragm: VITON
®
, 1.25” diameter (32mm)
•
Cable: TPE, 18 AWG - 2 wire shielded with vent tube, 0.3” diameter (8mm)
•
Stabilizing Weight: Epoxy coated cast iron, 1.0 lb. (0.45 kg),