PrimeLog
PrimeLog User Manual
13
If MEP interface is selected, then there is a choice of either single point or multiple point
calibration. For multi-point calibration, the given values or rate (metres/second) and
corresponding output frequencies (Hz) must be entered from the calibration certificate.
For both interfaces, single point calibration is simply entering the value given on the
calibration certificate.
PrimeWare automatically sets up the logger to calculate the flow rate in litres per second.
Please Note:
Quadrina devices must be configured with PrimeWare. They cannot be set up
using a Psion unit.
1.5.2 Other
Flowmeters
Interfacing the PrimeLog instruments to other flowmeter sensors requires knowledge of the
powering requirements of these sensors. The following table specifies the power interface:
Powering Standard Sensors
Type
Sensor
Power
source
1 -
Contact
Closure
Kent PSM/LRP/MSM/PU10
Kent Helix 4000 Reed Switch
Meinecke/Socam K01/K505R/K510/RO2.2/RD022
Schlumberger Model S/Cyble LF/Cyble HF
Datalogger
2 Kent
HRP/PU100
Datalogger
3 Kent
BPG20
Meinecke/Socam OPTO06/OPTO OD06
Schlumberger TLOS
Auxiliary
unit
4
ABB Kent-Taylor Aquamag/Aquaprobe
ABB Kent-Taylor Magmaster
Self-
powered
5
Kent Combination Meter
Datalogger
6
Combination Meter - non-Kent
Auxiliary
unit
Typical figures for the life of the auxiliary battery unit powering Type 3 sensors are:
Kent BPG20:
94 days
Socam OPTO06:
215 days
Schlumberger TLOS:
9 days