Introduction
Use this Quick Start Guide to quickly set up and understand the major
components of the
SmartXcan Mobile Pedestal
. It is critical to review these
materials to ensure proper usage of the device.
For further details, refer to the
.
Environmental Conditions
The SmartXcan requires specific environmental conditions for accurate
temperature readings. Use in a space with controlled environmental
conditions.
+
Ideal 64 - 75°F / 18 - 24°C
+
Relative non-condensing humidity 10-75%
+
Do not place next to a heat source
+
The thermal image sensor should not be aimed at
any other temperature source such as a incandescent
light source, radiator, or A/C system
+
Do not point at the sun or open sky
Acclimation (Warmup) Period
The Unit requires one minute to start up
and an estimated acclimation
period of 15 minutes. This acclimation period allows the temperature
scanning instrumentation to neutralize to the environment. The SmartXcan is
set to allow temperature scans before the acclimation period is completed,
but temperature accuracy can be affected.
If the unit is moved after the initial acclimation period, the device should be
allowed to re-acclimate for at least 15 minutes before performing additional
temperature measurements.
SmartXcan Mobile Pedestal
Accessories
Power Cable (IEC C13)
Varies based on country of use
Battery, 14.8V 98Wh
See Batteries Spec on page 7.
Battery Charger, Travel Verison
See User Manual for charging
information.
Battery Compartment Key (x2)
Indicator Light
Temperature Scanner