PRESYS
|
Instruments
DMY-2011, DMY-2035 and DMY-2036
Installation
Page 8
Attachment screw
Warning: follow indicated directions
when mounting rails..
Figure 4 - Dimensional drawing, panel mounting cutout and side view
for DMY-2011 Indicator
2.2 - Electrical Installation
DMY - 2011, 2035 and 2036 Indicators may be powered by voltage between
90 and 240VAC or 130 to 340VDC, any polarity. Remember that the internal circuit is
powered whenever the instrument is connected to the mains.
Input and output signals must be connected to the instrument only when it is
turned off.
Figure 5 shows the instrument rear terminals for connection to power supply,
ground, communication, process input and output signals.
Signal wiring must be kept far away from power wires.
Due to its metal case the instrument ground should be connected to earth
ground. Never connect the ground to neutral terminal.
presys