Thank you for choosing the Prestige Medical 
210047/210048 Classic Media Autoclave, designed to 
sterilize culture media. After removing from the box, please 
check for any transit damage. If any damage to the unit is 
found, please contact your supplier immediately.

Together with this unit and operating manual,  
you will find the following: 

  Electrical mains cord 

  Instrument furniture

  Spare fuses x2 (when fitted)

  Performance test certificate

  Certificate of Compliance (230v models only)

  Warranty registration card

The following descriptions refer to the pictures of the 
controls, display lights and operating symbols opposite.

Controls:

A

  Depressurization Valve

B

  Start Cycle Button

C

  Pressure Rise Indicator

D

  Interlock Mechanism

Display lights:

L1

 Power “On” light 

illuminates GREEN

L2

 Heating light 

illuminates ORANGE

L3

 Sterilizing light 

illuminates YELLOW

L4

 Sterilizing complete light 

illuminates GREEN

L5

 Fault light 

illuminates RED

Figures:

F1

 Lid aligned with body, turn clockwise to close

F2

 Lid in closed position

Warning symbols:

W1

  Caution, electric shock hazard

W2

  Warning, read manual before using the autoclave

W3

  Warning, unit must be earthed/grounded

W4

  Warning, heat hazard

Section 1:

 

Introduction

Section 2:

 

Operating symbols, displays and controls

WARNING!

Do not touch the body or lid as these parts become  

hot when the autoclave is in operation.

C

D

A

L2

L3

L4 L5

B

L1

F1

F2

W4

W3

W2

W1

3

Summary of Contents for Classic Media 210047

Page 1: ...211846B Classic Media User Manual Portable Steam Sterilizer...

Page 2: ...e of doubt contact Prestige Medical 0086 Prestige Medical Limited East House Duttons Way Shadsworth Business Park Blackburn BB1 2QR Tel 44 0 1254 682 622 Fax 44 0 1254 682 606 www prestigemedical co u...

Page 3: ...erating symbols opposite Controls A Depressurization Valve B Start Cycle Button C Pressure Rise Indicator D Interlock Mechanism Display lights L1 Power On light illuminates GREEN L2 Heating light illu...

Page 4: ...lease the interlock button Close the Depressurization Valve A 3 5 Power on Switch the unit on at the mains Light L1 illuminates GREEN 3 6 Temperature The 121 C temperature is automatically selected wh...

Page 5: ...ressurization Valve A in the open position when placing the lid upside down on a work surface 7 immerse the unit or electrical cord in water when cleaning 8 use abrasive materials or lubricants when c...

Page 6: ...r arrange for a service engineer to visit Fault 3 Sterilization failed to be achieved Light L4 fails to illuminate GREEN and there is no audible buzzer Disconnect from mains then reconnect and repeat...

Page 7: ...13A to BS1362 Rating Models are rated continuously for intermittent use Body Deep drawn aluminium Lid Aluminium Heater Externally surface mounted mechanically fixed electric element Temperature cut ou...

Page 8: ...of equipment contamination status must be available with the product Details of a suitable procedure are available on request Cleaning materials Mild washing up liquid Non abrasive cream cleaner Warm...