background image

 

                                                                                                                                                                                                   Page 11 of 19 

3

 

General operation 

3.1

 

Operation of the N-Gen nitrogen generator 

The N-Gen nitrogen generator regulates the pressure in a pressure vessel or directly to a consumer. How the N-Gen does this is 
explained in the next steps. The system requires a compressed air supply for this. 
 

3.1.1

 

Compressed air supply 

Step 1.

 

The compressor draws in the air from the environment and compresses it to a pressure of 10 bar(g).   Depending on the 
installation, this can also be lower with a minimum supply pressure of 6 bar. The compressor has a storage tank for 
compressed air and maintains constant pressure. From the storage vessel, the compressed air is transported to the 
nitrogen generator via a hose or a fixed line. In the nitrogen generator, the water and the oil residue are separated from 
the compressed air by two filters. 
 

3.1.2

 

Nitrogen production 

Step 1.

 

The pressure in the pressure vessel is monitored by means of a pressure switch or a pressure measurement from an 
external system. When the pressure becomes too low, a start command will be generated directly or indirectly and the N-
Gen will start producing the nitrogen. If there is no pressure vessel, the system is controlled by an external or other 
contact. 
 

Step 2.

 

Nitrogen production starts with drawing in compressed air from the compressor. The compressed air is transported to 
the PSA tubes where the Pressure Swing Adsorption process takes place. The N-Gen High Performance nitrogen 
generator has 2 PSA tubes for this process. The N-Gen Single Tube nitrogen generator uses only 1 PSA tube for this in 
conjunction with a nitrogen storage pressure vessel. The oxygen that is left over from the Pressure Swing Adsorption 
process is blown back into the immediate area, but can also be discharged into a channel or to the outside if required. 
The remaining mix of oxygen and nitrogen has a purity between 95.0 and 99.9999 vol .%N2. The nitrogen mix is fed to the 
pressure vessel and the pressure is increased. 
 

Step 3.

 

When the pressure is at the desired level, the start command is canceled. The nitrogen generator completes the 
production cycle. The compressor will continuously maintain the compressed air pressure. When compressed air is no 
longer used by the nitrogen generator, the compressor will also turn off. 
 

 

Summary of Contents for N-Gen

Page 1: ...N Gen nitrogen generator High Performance and Single tube User manual Type User manual Date March 2021 Version Revision Version 2 0 Revision 00 Status Approved Company PRESSCON Location Honselersdijk...

Page 2: ...Page 2 of 19...

Page 3: ...1 OPERATION OF THE N GEN NITROGEN GENERATOR 11 3 1 1 Compressed air supply 11 3 1 2 Nitrogen production 11 3 2 SCHEMATIC OVERVIEW OF THE N GEN NITROGEN GENERATOR 12 4 CONTROLLER 13 4 1 MULTI F CONTROL...

Page 4: ...rence is in the number of PSA tubes that are used to make nitrogen The High Performance version uses 2 PSA tubes and the Single tube version uses only 1 PSA tube The Single Tube generator is a simplif...

Page 5: ...tly from the output connection may cause intoxication The nitrogen quickly mixes with the ambient air so that at a distance of 0 5m from the nitrogen output connection it no longer poses a health risk...

Page 6: ...s with bare hands Use heat resistant clothing 3 DANGEROUS TO INHALE GAS Avoid the specified area or wear appropriate breathing equipment 9 MOVING PARTS Risk of getting stuck or injuries to limbs Turn...

Page 7: ...ents by providing an illustration and numbering the most important parts The compressor is not described because different brands can be used that come with their own manual 2 1 N Gen nitrogen generat...

Page 8: ...e 8 of 19 2 2 N Gen nitrogen generator High Performance H P N gen controller 2 2 14 7 FX Pre ve nt 12 1 3 3 5 13 6 11 8 4 Figure 2 Schematic illustration of the N Gen High Performance nitrogen generat...

Page 9: ...Page 9 of 19 2 2 14 7 FX Pre ve nt 12 1 3 3 5 13 11 8 4 8 8 6 15 8 16 Figure 3 Schematic illustration of the N Gen High Performance nitrogen generator...

Page 10: ...7 Compressed air inlet Push in connection that is used for connecting the compressor to the nitrogen generator This is commonly done with a 16mm nylon air hose 8 PSA control manifold A pneumatically c...

Page 11: ...mand will be generated directly or indirectly and the N Gen will start producing the nitrogen If there is no pressure vessel the system is controlled by an external or other contact Step 2 Nitrogen pr...

Page 12: ...mponents in the system communicate with each other N Gen Control Nitrogen generator Compressor Filter PSA valve 1 PSA valve 2 PSA valve 3 Nitrogenpressure OK NO Alarm Air pressure to low NO Drainvalve...

Page 13: ...board for measuring and controlling A N Gen nitrogen generator consists of one or more N Gen Controls For the N Gen Control two hardware versions can be used the JGA2000 and the JGA2024 These boards a...

Page 14: ...The N Gen nitrogen generator therefore has two alarm levels If one of the alarms occurs the nitrogen generator is stopped and the main valve and the drain valve are closed The nitrogen generator is s...

Page 15: ...d out the controller Please contact PRESSCON Also in case of technical failure questions or other issues you can contact PRESSCON at any time 5 2 N Gen control 5 2 1 Faults and alarms The error levels...

Page 16: ...Check compressed air filters Service engineer X Check output flow and purity Service engineer X Clean dirt catcher of compressor Service engineer X Replace the compressed air filters Service engineer...

Page 17: ...negligence accident or whatsoever The warranty includes checking repair or replacement if defective from proven equipment Faulty equipment which has been replaced shall become property of PRESSCON Reg...

Page 18: ...has been added to the document Revision _R01 to _R02 indicates Correction an image language error or text error has been corrected Version Date Modification 1 1_R00 20 04 2020 Manual version 1 1_R00 t...

Page 19: ...rking with a product of PRESSCON Its aim is to maintain and possibly improve the quality of its products If additional information or support is requested then PRESSCON can be consulted through the fo...

Reviews: