background image

9

User Manual

All illustrations in this manual are for information purposes only and may differ from the product supplied by us.

Philips TUV TL Mini

General information

Approval and application

Cap-Base

G5  [G5] 

Mercury (Hg) Content (Nom)

4.4 mg

Main Application

Disinfection 

Useful Life (Nom)

6000 h 

UV

System Description

-

UV-C Radiation

0.9 W

Light technical

Product data

Color Code 

TUV

Full product code

871150063872427

Color Designation

- [Not Specified]

Order product name

TUV 4W FAM/10X25BOX

Depreciation at Useful Lifetime

0,25%

EAN/UPC - Product

8711500638724

Order code

928000104013

Operating and electrical

Numerator - Quantity Per Pack

1

Power (Nom)

4,5 W

Numerator - Packs per outer box

250

Lamp Current (Nom)

0.17 A

Material Nr. (12NC) 

928000104013

Voltage (Nom)

29 V

Net Weight (Piece) 

18.600 g

Mechanical and housing

Cap-Base Information

2 Pins

Bulb Shape

T16

Product specifications

Warnings and safety

It is very unlikely that the breaking of a lamp will have any impact on your health. If a lamp breaks, ventilate the room for 30 
minutes and remove the parts, preferably using gloves. Place them in a sealed plastic bag and take the bag to your local 
waste disposal company for recycling. Do not use a vacuum cleaner. 

HAZARD:

Product of risk group 3 Ultra Violet. These lamps emit powerful UV radiation that can cause serious injury to the skin and 
eyes. Avoid exposure of eyes and skin to unshielded products. Use only in an enclosed environment that protects users 
from the radiation.

Summary of Contents for Cart UV-C

Page 1: ...Predia Cart UV C User Manual...

Page 2: ...are for information purposes only and may differ from the product supplied by us Table of contents Predia Cart UV C Control Panel 3 Description of the three switching modes 4 Setting the UV switch and...

Page 3: ...ages The control panel is located on the back of the Predia Cart UV C and can be accessed by removing the large grey back cover After setting up the Predia Cart UV C you can switch on the charging car...

Page 4: ...ect this mode you switch on all the power sockets They will only turn off when you press the on off button on the front of the Predia Cart UV C Plug the power cable supplied into the power socket and...

Page 5: ...switch must be switched on every day by means of the on off button on the front of the Predia Cart UV C Plug the power cable supplied into the power socket and connect the C13 plug device plug to the...

Page 6: ...of the on off button on the front of the Predia Cart UV C Plug the power cable supplied into the power socket and connect the C13 plug device plug to the side of the cart Slide the on off switch to t...

Page 7: ...mode when the Switch is pressed at the moment the device is switched on After that you can set the time or adjust the mode 3 You can select the control function by setting the switch on Mode options...

Page 8: ...tion the UV C lamps are automatically turned off Set disinfection time The disinfection time is set to 30 minutes as standard and in practice this is more than sufficient However if you would prefer t...

Page 9: ...ower Nom 4 5 W Numerator Packs per outer box 250 Lamp Current Nom 0 17 A Material Nr 12NC 928000104013 Voltage Nom 29 V Net Weight Piece 18 600 g Mechanical and housing Cap Base Information 2 Pins Bul...

Page 10: ...manual are for information purposes only and may differ from the product supplied by us A B C D Philips TUV TL Mini TUV 4W Dimensial drawing 5 Product ID D B B C A TUV 4W FAM 10X25BOX 16 mm 143 mm 140...

Reviews: