OLT-301 Series 

Safety Instructions 

Safety Terms Used in This Manual 

 

WARNING 

The WARNING sign denotes a hazard. It calls attention to a procedure which could result in personnel injury. 

 

CAUTION 

The CAUTION sign denotes a hazard. It calls attention to an operating procedure if not correctly performed or 
adhered to, could result in damage to or destruction of part or the entire product. 

 

NOTE 

The NOTE sign information that may be beneficial during the use and maintenance of the instrument. 

 

WARNING 

OLT-301 Series Optical Loss Tester is a laser device. User should always avoid direct eye exposure to the laser 
output. Using microscope or magnifier to observe the laser output should also be avoided: laser beam may 
converge on the retina and cause permanent eye injury. 

 

CAUTION 

Battery

: OLT-301 Series battery type is lithium battery. 

Battery Power Supply

: Do not expose battery to fire or intense heat. Do not open or mutilate battery. Avoid 

touching the electrolyte in the battery, which is corrosive and may cause injuries to eyes, skin or damage to 
clothes. 

External Power Supply

: OLT-301 Series support external power supply: 5V DC/750mA. External power supply 

is optional. 

Laser Radiation

: To avoid serious eye injury, never look directly into the optical outputs of fiber optic network 

equipment, test equipment. 

 

Always avoid looking directly into the optical output port, when the instrument is working. 

 

Always replace protective dust cap on the detector port when the instrument is not in use. 

 

Always avoid looking directly at unconnected end of optic fiber in testing and make the unconnected end 
pointing at a non-reflective object. 

 

 

Summary of Contents for OLT-301 SERIES

Page 1: ...OLT 301 SERIES Optical Loss Tester Operation Guide V 1 23 17...

Page 2: ...2 3 Indicator 8 2 3 Use of Battery 8 2 4 Connector Cleaning 8 3 Operation 9 3 1 Power On 9 3 2 Stabilized Laser Source SLS 11 3 2 1 SLS Laser On Off 12 3 2 2 SLS Reference Setting 12 3 2 3 SLS Setup 1...

Page 3: ...RL Test 22 3 5 2 ORL Save Record 23 3 5 3 ORL Recall 24 3 5 3 ORT Setup 24 4 Maintenance and Calibration 25 4 1 Cleaning of the Connectors 25 4 2 Calibration Requirement 25 5 Warranty Information 25 5...

Page 4: ...voided laser beam may converge on the retina and cause permanent eye injury CAUTION Battery OLT 301 Series battery type is lithium battery Battery Power Supply Do not expose battery to fire or intense...

Page 5: ...o the shipping agent or the representative of PRO immediately Retain the original packing materials in case reshipment is necessary 1 2 Unpacking and Inspection This instrument has been carefully pack...

Page 6: ...o test records save in local unit remote unit both units USB power charging Over 35 hours continuous operation Multilanguage support No warm up quick start High resolution color LCD Pocketsize lightwe...

Page 7: ...cal Input Output The OLT 301 Series is available with FC connector Interchangeable SC ST Note Output port is marked with LS Input port is marked with PM 2 USB Power Data Port The OLT 301 Series can be...

Page 8: ...w the instructions below when cleaning Turn off the instrument before cleaning Non compliant operation may result in hazardous radiation exposure Turn off laser source before cleaning optical interfac...

Page 9: ...Operation 3 1 Power On Press and loading screen appears Then it enters functional interface The last interface before last shutdown see Figure 3 2 Figure 3 2 The toggle between functional interfaces P...

Page 10: ...ess to enter the corresponding function interface The toggle between different setting interfaces Under function interface press and to toggle the following interfaces Under function interface press a...

Page 11: ...ess and to select Mass storage PC recognition it for a USB flash drive Communication the equipment directly transfer the data to the PC software mainly used in the optical power meter Draw Trance func...

Page 12: ...etween dBm and mW 3 2 2 SLS Reference Setting Note This function is only for AutoID mode applicable Ref XXX XXdBm parameters will set up to the opposite side which has the corresponding models test in...

Page 13: ...engths 3 2 3 SLS Setup Press and to Setup interface see Figure 3 7 Note This function is only for Auto ID mode applicable Figure 3 7 While the cursor on Auto ID press to select Manual or Automatic Und...

Page 14: ...ce Precision Rated Optics laser source with AutoID function on the other side of the tested fiber enables AutoID the OPM will automatically set the reference value sent from the laser source as refere...

Page 15: ...nterface see Figure 3 10 Figure 3 10 Press and to select item Del all Del current Go to Index Prev and Next 3 3 3 OPM Trace Press and to Trace sub interface see Figure 3 11 Figure 3 11 Press and to se...

Page 16: ...nd 8hour Status press to running stop trace 3 3 4 OPM Setup Press and to Setup sub interface see Figure 3 13 Figure 3 13 Press and to select item Zero and Unit Zero Screw on the dust cap press the twi...

Page 17: ...ue 5 Save 6 Toggle the wavelength and then repeat step 3 4 and 5 3 4 Optical Loss Test OLT Press and to Optical Loss Test interface see Figure 3 14 Figure 3 14 3 4 1 OLT Reference Setting Press and to...

Page 18: ...easonable which is recommended After connecting the ports press and to select Start press to start the test Press and to select Save press to save the measured value as reference value 3 4 2 OLT Bi di...

Page 19: ...gure 3 17 Figure 3 17 Press and to select item Cable and Fiber press to enable digit adjustment press to confirm Press and to select Save and press to save current record 3 4 4 OLT Recall Press and to...

Page 20: ...lt PASS or FAIL according to the thresholds setting Loss dB km In bi directional test the unit will calculate the fiber s approximate length according to the attenuation dB km set here Please note tha...

Page 21: ...ly calibrate zero 2 Reference Setting Ref Type set to twin 3 Threshold Setting set Thres and Loss the value of the Thres limit the maximum loss of the entire link Loss is generally provided by the fib...

Page 22: ...Return Loss Test ORL Option Press to Optical Return Loss sub interface as shown in Figure 3 22 Figure 3 22 3 5 1 ORL ORL Test Press and to Test sub interface as shown in Figure 3 22 Press and to selec...

Page 23: ...t return loss value as the ORL Zero ORL Ref 14 7dB 14 7dB means the current ORL value measured under default ORL zero value and default ORL reference Connect the reference patch cord to OLT port press...

Page 24: ...sub interface as shown in Figure 3 24 Figure 3 24 Press and to select item Del all Del current Go to Index Prev and Next 3 5 3 ORT Setup Press and to Setup sub interface as shown in Figure 3 25 Figur...

Page 25: ...Keep the cleanness of connectors and dust caps The detector needs to be cleaned timely 4 2 Calibration Requirement Calibration of OLT 301 Series is recommended every three years Please contact PRO Te...

Page 26: ...n Materials Authorization number And describe briefly reasons for the return of the equipment to allow us offer you more efficient service NOTE To return the instrument in the case of repair calibrati...

Page 27: ......

Page 28: ...on Rated Optics Inc Corporate Office Billing Processing PO Box 877 Trexlertown PA 18087 Precision Rated Optics Inc Product Distribution Center Manufacturing Testing 9999 Hamilton Blvd Breinigsville PA...