OLT-301 Series
3.4.5 OLT-Threshold Setting
Press
and
to [Thresholds] sub-interface, as shown in Figure 3.18.
Figure 3.18
Press
and
to select the item to be adjusted, press
to enable adjustment, and press
to
confirm.
NOTE
Thres:
In bi-directional test, the OLT-301 Series determines test result
PASS or FAIL
according to the
thresholds setting.
Loss:
dB/km
, In bi-directional test, the unit will calculate the fiber’s approximate length according to the
attenuation (dB/km) set here. Please note that the attenuation at each wavelength is setup separately, hence the
calculated length may be different at each wavelength. Non-fiber attenuation factors (connector, splices) and
incorrect attenuation setting may also affect the length calculation.