8
INSTALLATION
WARNING
• Read this User Manual before installing or operating the Oxygen
Conserving Regulator.
• Inspect the Conserver and cylinder valve to ensure they are free of
oils, greases or other contaminants.
• DO NOT
direct flow of oxygen at any person, or flammable material
when cracking open the cylinder.
• DO NOT
use without yoke seal.
• DO NOT
use plastic yoke seals.
• Use Precision Medical, Inc. yoke seal made of viton and brass.
• The yoke seal supplied with the Conserver is reusable.
CAUTION
• Inspect the Oxygen Conserving Regulator for visual damage before
use, DO NOT USE if damaged.
• DO NOT
use tools to tighten T-handle, this can lead to over tightening
and will cause damage to the Conserver.
• Be sure all connections are tight and leak free.
• DO NOT
use liquid leak detector to test for leaks.
Note
: For Operational Instructions on how to operate the cylinder
consult with your Equipment Provider.
1. Position the cylinder so the oxygen cylinder valve outlet is pointing
away from the user and any other person(s).
2. Before installing the Conserver, “crack” cylinder valve by opening
the valve slightly to remove any foreign particles and then
close valve.
3. Be sure the Conserver is in the “OFF” position. Align “OFF” with the
indicating pointer .
4. Be sure the CGA 870 connection high pressure inlet has a
yoke seal.
5. Align Index Pins with Index Pin Alignment Holes in Post Valve.
6. Hand tighten T-handle until snug.