Precision Medical EasyPulse 5 User Manual Download Page 5

4

 WARNING

•  DO NOT USE

 without a seal. For optimum safety, use only a Precision Medical, Inc. Inlet 

Seal. The Inlet Seal supplied with the Conserver is reusable. 

DO NOT USE

 any other 

type of seal. 

•  To ensure receiving the therapeutic amount of oxygen delivery according to your medical 

condition the EasyPulse Conserver must; 

  •  be used only after one or more settings have been individually determined or    

    prescribed for you at your specific activity levels

 

  •  be used with the specific combination of parts and accessories that are in line with 

     the specification of the oxygen conserver manufacturer and that were used while  

    your settings were determined.

•  The pulse settings of the EasyPulse Conserver might not correspond with continuous  

flow oxygen.

•  The settings of other models or brands of oxygen therapy equipment do not correspond 

with the settings of the EasyPulse Conserver.

•  Use only water based lotions or salves that are oxygen compatible during setup or use 

during oxygen therapy. Never use petroleum or oil-based lotions or salves to avoid the 

risk of fire and burns. Wash and dry hands properly prior to use.

• 

Do not lubricate replaceable fittings, connections, tubing, or other accessories of the 

oxygen conserver to avoid the risk of fire and burns.

•  Use only spare parts recommended by the manufacturer to ensure proper function and to 

avoid the risk of fire and burns.

•  Wind or strong draughts can adversely affect accurate delivery of oxygen therapy. 

Using this equipment beside an open window or in front of a fan can affect the accuracy 

of delivery of oxygen. 

Using this equipment in the back seat of an open convertible car can affect the accuracy 

of delivery of oxygen.

•  Use of this device at an altitude above 10,000 ft (3048 m) or outside a temperature of 

35°F to 105°F (1.7°C to 40.6°C) is expected to adversely affect the quality of the therapy.

• 

Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula on 

bed coverings or chair cushions, if the oxygen conserver is turned on, but not in use; the 

oxygen will make the materials flammable. Turn the oxygen conserver off when 

 

not in use.

•  If you feel discomfort or are experiencing a medical emergency, seek medical assistance 

immediately to avoid harm.

•  Geriatric, pediatric or any other patient unable to communicate discomfort can require 

additional monitoring to avoid harm.”

•  Smoking during oxygen therapy is dangerous and is likely to result in facial burns 

or death. Do not allow smoking or open flames within the same room of the oxygen 

conserver or any oxygen carrying accessories. If you smoke, you must always turn the 

oxygen conserver off, remove the cannula and leave the room where either the cannula 

or the oxygen conserver is located. If unable to leave the room, you must wait 10 minutes 

after the flow of oxygen has been stopped.

Summary of Contents for EasyPulse 5

Page 1: ...62 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485 Certified www precisionmedical com SAVE THESE INSTRUCTIONS 198705 shown Federal USA law restricts this device to sale by or on the order...

Page 2: ...ARATION OF CONFORMITY 15 RECEIVING INSPECTION Remove the Precision Medical Inc EasyPulse5 Oxygen Conserving Regulator Conserver from the packaging and inspect for damage If there is any damage DO NOT...

Page 3: ...minently hazardous situation which if not avoided will result in death or serious injury WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injur...

Page 4: ...Conserver Wash and dry hands properly prior to use DO NOT use a humidifier with the Conserver DO NOT allow cylinders to tip or fall Secure gas cylinders so they cannot fall For optimum safety keep cy...

Page 5: ...nsure proper function and to avoid the risk of fire and burns Wind or strong draughts can adversely affect accurate delivery of oxygen therapy Using this equipment beside an open window or in front of...

Page 6: ...ys The flow is determined by setting the flow control knob to the prescribed flow The oxygen is supplied to the patient through the cannula CAUTION Only personnel instructed and trained in its use sho...

Page 7: ...uency Up to 35 b min Cannula Requirement Maximum 7 foot 2 1 m long standard adult single lumen nasal cannula Operating Conditions Temperature 35 F to 105 F 1 7 C to 40 6 C Altitude Sea level to 10 000...

Page 8: ...OMPONENT DESCRIPTION CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc Index Pins Yoke Seal Cannula Connection T Handle Dial Contents Gauge Indicator Pointer 198705 sh...

Page 9: ...to tighten T handle this can lead to over tightening and will cause damage to the Conserver Be sure all connections are tight and leak free DO NOT use liquid leak detector to test for leaks Note For O...

Page 10: ...quipment 2 Place the cannula over your ears and position the prongs in your nose as instructed by your Provider or cannula manufacturer 3 Slowly open cylinder valve counterclockwise until completely o...

Page 11: ...kPa and below it is recommended to change to a full oxygen cylinder NO OXYGEN is delivered in between settings WARNING Use only Precision Medical Inc carry bag designed for the EasyPulse5 to prevent...

Page 12: ...ackaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in transit Refer to Precision Medical Inc Return Policy available on the Internet www pre...

Page 13: ...in OFF position 3 Cylinder empty 4 Dial not set at proper position 5 Conserver not sensing breath 1 Open the valve on cylinder 2 Set to prescribed setting 3 Replace cylinder 4 Set dial so indicator po...

Page 14: ...TS Description Part Cannula 504833 Yoke Seal 1221 Carry Bag with shoulder strap Conserver properly installed in Carry Bag ACCESSORIES Description Part Carry Bag M6 503920 Carry Bag M4 M6 504184 Carry...

Page 15: ...make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete a...

Page 16: ...of Conformity see our website http www precisionmedical com documentation Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherla...

Reviews: