Precision Medical EasyPulse 5 User Manual Download Page 7

6

SPECIFICATIONS

Inlet Pressure Range:

300 - 3000 psi (2068 - 20684 kPa)

Pressure Gauge Accuracy:

3 - 2 - 3% of full scale

Dimensions: 

(Are approximate and may vary by model)

 

Model: 198705

 Weight:

9.9 oz 

(279 g)

 

Overall Length:

4.75 in  (12.07 cm)

 Width:

2.63 in  (6.67 cm)

 Height:

2.50 in  (6.35 cm)

Pulse Settings:

1, 2, 3, 4, & 5 l/min Equivalents

Flow Settings:

2 l/min Continuous (198705)

1 through 6 l/min Continuous (198712)

Accuracy: 

Continuous Settings:

within ± 10%

PULSE ONL

Y Accuracy:

Pulse Settings: /- 15% of the nominal 

bolus value (at each breath rate)

Savings Ratio:

Up to 5.7:1

Trigger Method:

Inspiratory effort (Negative pressure from  

patient inhalation)

Breathing Frequency:

Up to 35 b/min

Cannula Requirement:

Maximum 7 foot (2.1 m) long standard adult 

single lumen nasal cannula.

Operating Conditions:

 Temperature:

35°F to 105°F (1.7°C to 40.6°C)

 Altitude:

Sea level to 10,000 ft (0 to 3,048 m)

Storage Conditions:

 Temperature:

-40°F to 140°F (-40°C to 60°C)

  Maximum Humidity:

95% Noncondensing

Oxygen Cylinder Connection:

CGA 540 Valve

Ignition and Fault Tolerance:

Meets ASTM G175-03

Specifications are subject to change without prior notice. Manuals available on our Website; www.precisionmedical.com
Specifications for the flowrates and volume for the gas delivered to the patient are expressed in STPD (standard temperature and 

pressure dry) conditions.(14.69 psi (101,3 kPa) at an operating temperature of 68 °F (20 °C)), dry

Specification tolerances include the uncertainty of the measurement used to determine the specification.

Summary of Contents for EasyPulse 5

Page 1: ...62 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485 Certified www precisionmedical com SAVE THESE INSTRUCTIONS 198705 shown Federal USA law restricts this device to sale by or on the order...

Page 2: ...ARATION OF CONFORMITY 15 RECEIVING INSPECTION Remove the Precision Medical Inc EasyPulse5 Oxygen Conserving Regulator Conserver from the packaging and inspect for damage If there is any damage DO NOT...

Page 3: ...minently hazardous situation which if not avoided will result in death or serious injury WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injur...

Page 4: ...Conserver Wash and dry hands properly prior to use DO NOT use a humidifier with the Conserver DO NOT allow cylinders to tip or fall Secure gas cylinders so they cannot fall For optimum safety keep cy...

Page 5: ...nsure proper function and to avoid the risk of fire and burns Wind or strong draughts can adversely affect accurate delivery of oxygen therapy Using this equipment beside an open window or in front of...

Page 6: ...ys The flow is determined by setting the flow control knob to the prescribed flow The oxygen is supplied to the patient through the cannula CAUTION Only personnel instructed and trained in its use sho...

Page 7: ...uency Up to 35 b min Cannula Requirement Maximum 7 foot 2 1 m long standard adult single lumen nasal cannula Operating Conditions Temperature 35 F to 105 F 1 7 C to 40 6 C Altitude Sea level to 10 000...

Page 8: ...OMPONENT DESCRIPTION CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc Index Pins Yoke Seal Cannula Connection T Handle Dial Contents Gauge Indicator Pointer 198705 sh...

Page 9: ...to tighten T handle this can lead to over tightening and will cause damage to the Conserver Be sure all connections are tight and leak free DO NOT use liquid leak detector to test for leaks Note For O...

Page 10: ...quipment 2 Place the cannula over your ears and position the prongs in your nose as instructed by your Provider or cannula manufacturer 3 Slowly open cylinder valve counterclockwise until completely o...

Page 11: ...kPa and below it is recommended to change to a full oxygen cylinder NO OXYGEN is delivered in between settings WARNING Use only Precision Medical Inc carry bag designed for the EasyPulse5 to prevent...

Page 12: ...ackaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in transit Refer to Precision Medical Inc Return Policy available on the Internet www pre...

Page 13: ...in OFF position 3 Cylinder empty 4 Dial not set at proper position 5 Conserver not sensing breath 1 Open the valve on cylinder 2 Set to prescribed setting 3 Replace cylinder 4 Set dial so indicator po...

Page 14: ...TS Description Part Cannula 504833 Yoke Seal 1221 Carry Bag with shoulder strap Conserver properly installed in Carry Bag ACCESSORIES Description Part Carry Bag M6 503920 Carry Bag M4 M6 504184 Carry...

Page 15: ...make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete a...

Page 16: ...of Conformity see our website http www precisionmedical com documentation Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherla...

Reviews: