P
RO
V
U
™ PD6363 Dual Analog Input Rate/Totalizer
Instruction Manual
33
Setting the Display Parameters &
Intensity (
dsplay
)
Display line 1 (
line 1
) can be programmed to
display:
1. Ch-A rate (
D Ch-A
)
2. Ch-B rate (
D Ch-B
)
3. Ch-C math channel (
D Ch-C
)
4. Toggle Ch-A & Ch-B rate (
D AB
)
5. Toggle Ch-A rate and Ch-C (
D AC
)
6. Toggle Ch-B rate and Ch-C (
D BC
)
7. Toggle Ch-A & Ch-B rate, and Ch-C (
D ABC
)
8. Ch-A total (
D T-A
)
9. Ch-B total (
D T-B
)
10. Ch-A grand total (
D GT-A
)
11. Ch-B grand total (
D GT-B
)
12. Toggle Ch-A rate and total (
D RT-A
)
13. Toggle Ch-B rate and total (
D RT-B
)
14. Toggle Ch-A rate and grand total (
DRGT-A
)
15. Toggle Ch-B rate and grand total (
DRGT-B
)
16. Relay set points (1-8) (
DSET 1
to
DSET 8
)
17. Max, min, and max & min values for
Ch-A, Ch-B, or Ch-C (
D Hi-A
to
d HL-C
)
18. Toggle Ch-A rate & units (
d A-u
)
19. Toggle Ch-B rate & units (
d B-u
)
20. Toggle Ch-C & units (
d C-u
)
21. Toggle Ch-A total & units (
d TA-u
)
22. Toggle Ch-B total & units (
d TB-u
)
23. Toggle Ch-A total and Ch-B total (
d tAb
)
24. Toggle Ch-A total, Ch-B total, and the sum of
total A + total B (
d tABC
)*
25. Modbus input (
nm bus
)
*Note: The (C = sum of total A + total B) displayed
with the selection
(
d tABC
)
is different than the
math channel C calculated under the Math menu
functions. Example: (C = Rate Ch-A + Rate Ch-B).
Display line 2 (
Line 2
) can be programmed to
display:
1. All options for display line 1
2. Off, with no display (
d ofF
)
3. Engineering units for any single channel,
total, or grand total
Display Parameter Menu
After setting up the input and display, press the Menu
button to exit programming and skip the rest of the
setup menu.
Customizable Displays
The displays can be set up to read input channels
(A or B), rate, total, or grand total, math function
channel C, toggle between A & B, B & C, A & C, A &
B & C, toggle between channels A, B, or C & units,
the max/min of any of the channels, including the
math channel (C), set points, or the Modbus input.
This allows the display to be setup to display
whatever variables are most valuable to the
application. Here are just a few examples.
Display Intensity (
d-Inty
)
The meter has eight display intensity levels to give the
best performance under various lighting conditions.
Select intensity 8 for outdoor applications. The default
intensity is 6.