ProtEX-MAX PD8-6200 Explosion-Proof Analog Input Rate/Totalizer Instruction Manual
61
Parameter
Display Default Setting
Password 1
Pass 1
000000 (unlocked)
Password 2
Pass 2
000000 (unlocked)
Password 3
Pass 3
000000 (unlocked)
Parameter
Display Default Setting
Total password
Total
000000 (unlocked)
Grand total
password
Gtotal
000000 (unlocked)
Troubleshooting Tips
Symptom
Check/Action
SafeTouch buttons do not
respond
If mechanical button was pushed. The SafeTouch buttons will be re-enabled
automatically
60 seconds
after the last button push.
If slide switch on connector board is in DISABLE position, switch to ENABLE.
Strong direct sunlight may interfere with SafeTouch button operation. It is
recommended to operate the buttons by standing so as to block direct
sunlight.
Serial Communications Power
LED Indicator is off
1. Check modular cable connection
Check power to the device
If only the TX (or DATA IN)
data status LED is flashing
when serial communications
attempted
1. Check serial cable
2. Check protocol selected on device
3. Check instrument address & baud rate
Check program address & baud rate
If both data status LEDs (TX
and RX) are off when trying to
communicate
Remove all unnecessary cables and instruments from the bus. Try getting
the system to work with only one device (to ease troubleshooting) and then
expand the system one device at a time.
Communications slow
Increase the baud rate
Random communication errors
1. Increase the TX delay time
Decrease the baud rate
Power LED is off
1. Check modular cable connection
Check power to instrument
No display at all
Check power at power connector
Not able to change setup or
programming,
Locd
is
displayed
Meter is password-protected, enter correct six-digit password to unlock
Meter displays error message
during calibration (
Error
)
Check:
1. Signal connections
2. Input selected in
Setup
menu
3. Minimum input span requirements
Meter displays
999999
-99999
Check:
1. Input selected in
Setup
menu
2. Corresponding signal at Signal connector
Display is unstable
Check:
1. Input signal stability and value
2. Display scaling vs. input signal
3. Filter and bypass values (increase)
Display response is too slow
Check filter and bypass values
Display reading is not accurate Check:
1. Signal input conditioner selected: Linear, square root, etc.
2. Scaling or calibration
Display does not respond to
input changes, reading a fixed
number
Check:
Display assignment, it might be displaying max, min, or set point.
Display alternates between
1.
Hi
and a number
2.
Lo
and a number
Press Menu to exit max/min display readings.